Sensor catalog - Watlow - #35

/ 420


catalogue search
P. 01
P. 02
P. 03
P. 04
P. 05
P. 06
P. 07
P. 08
P. 09
P. 10
P. 11
P. 12
P. 13
P. 14
P. 15
P. 16
P. 17
P. 18
P. 19
P. 20
P. 21
P. 22
P. 23
P. 24
P. 25
P. 26
P. 27
P. 28
P. 29
P. 30
P. 31
P. 32
P. 33
P. 34
P. 35
P. 36
P. 37
P. 38
P. 39
P. 40
P. 41
P. 42
P. 43
P. 44
P. 45
P. 46
P. 47
P. 48
P. 49
P. 50


See other catalogues for Watlow

Text version of the page
General Information
Lab Services
Quality Certification Lab
Continued
Certification Testing
Service
Description
Specifications
End-to-End Calibration
Comparison of each end of a length of thermocouple wire by utilizing a common junction measurement test. This is a requirement to verify homo­geneity requirements.
ASTM E 207, E 220, E 230
Dielectric Testing
Performance levels of wire insu­lations in the presence of high, local fields caused by electrical discharges. Routinely used in Watlow quality control testing.
ASTM D 149
Helium Leak
Test
Verifies the sheath integrity in metal-sheathed cable and sen­sors to 70 kg/cm2 (1000 psi) in specially designed pressure chambers.
ASTM E 235
Radiographic Inspection
Determines dimensions, and detects and evaluates cracks, voids, inclusions and discon­tinuities. Technicians qualified under SNT-TC-1A.
ASTM E 94, E 142,
Metallographic Examination
Reveals the constituents and structures of metals. Photomicrographs are also avail­able to determine and document average grain size and structure of prepared specimens.
ASTM E 3, E 112, E 235
Compaction Density Test
Determines the compaction of insulating materials in metal-sheathed cable.
ASTM D 2771
Drift Test
Determines long-term stability and drift characteristics.
ASTM E 601, E 644
Thermal Cycle Test
Individual sensors subjected to repeated cycling through a temperature range.
ASTM E 235
Insulation Resistance
Measures the electrical insu­lation resistance properties between the thermoelements and the sheath at ambient as well as elevated tempera­tures to determine the presence of moisture or impurities which could affect sensor performance.
ASTM E 780,
E 235, E 644
Spurious EMF
Determines the homogeneity of the thermoelements. Per­formed at high temperatures on the entire length of XACTPAK mineral insulated, metal-sheathed cable.
Watlow
Micro-Hardness
Determines the hardness of sheath or conductors used to measure a material's resistance to penetration (hardness) as a predictor of strength, machin-ability, brittleness, ductility and wear resistance.
Vicker's
(D 3 (D
a
3
k
o 3
0) -+
o
3
(D
i
c
(D
s
X-rays of the sensor verify the nonexistence of cracks at weld points that could let in humidity or gas and potentially shorten the life of the thermocouple.
35