In North America, call 1-972-446-9990 49
Detectors PM100
Single Wavelength PM140
PM100 Detector
PM140 Detector
PM100/140 Pin-Out Assignments
DB9M D-Subminiature Connector
Pin Description
1 1Remote high voltage programming, 2–10VDC
2 Power input line, –12 to –15VDC (1mA), polarity protected
3 Power input line, +12 to +15VDC (7mA), polarity protected
4 Not used
5 Signal output, 0–10VDC
6 Not used
7 Circuit Ground Return
8 Not used
9 Circuit Ground Return
1This enables automatic gain control when used with any of Verity’s
system controllers.
Features and Benefits
.. Cost-effective and compact through the use of
interference filters
.. Enables measurement in low light applications
through high-gain photomultiplier tube
.. Adjusts for varying light conditions through
automatic gain control feature when used with
Verity’s system controllers
.. Simple installation through a KF40 flange or
mounting bracket
Description
The PM100/140 Series detectors are used to measure a
single fixed wavelength in the range of 185–830nm.
When used with Verity’s system controllers, the
PM100/140 Series provides robust endpoint detection
when used in the appropriate applications. This series is
most commonly used in applications where the
chemistry is simple enough that a fixed 10nm bandpass
detector is suitable. At some wavelengths, alternative
filters with a lower bandpass may be available.
The PM100 mounts to a viewport and includes an
integral light aperture located on the bottom of the
instrument. The PM140 includes a KF40 flange for
mounting to the vacuum chamber. A quartz or sapphire
window provides vacuum isolation. A sapphire window
provides better resistance to window etching than the
quartz window.
|
|