You may also be interested in
Sensor, Cup anemometer, Digital indicator, Anemometer, Dew-point meter
Text version of the page
- If sampling through a hole or other port in the
incubator, insert Nafion® tubing into the incubator
and seal around it.
- If sampling through a hose barb fitting/sample
port, use a very short length of Neoprene tubing
to connect the fitting of the Nafion tubing to
the incubator fitting. Use the Neoprene as a
“connector” to hold the Nafion® tubing as close
as possible to the hose barb. It is undesirable to
have the sample gas travel through the Neoprene,
as condensation is likely to result within the
Neoprene tubing
- As a precaution, keep the GM70 pump above
the level of the chamber sampling port. If
condensation occurs in the sample line, this
will prevent liquid water from damaging the CO2
sensor.
For more information, visit
www.vaisala.com or contact
us at sales@vaisala.com
Ref. B210826EN-A ©Vaisala 2009
This material is subject to copyright protection, with all copyrights
retained by Vaisala and its individual partners. All rights reserved.
Any logos and/or product names are trademarks of Vaisala or
its individual partners. The reproduction, transfer, distribution or
storage of information contained in this brochure in any form
without the prior written consent of Vaisala is strictly prohibited.
All specifications — technical included — are subject to change
without notice.
Td (°C) -40 -30 -20 -10 0 10 20 30 40 50 60
-60 0.9999 0.9996 0.999 0.997 0.994 0.988 0.977 0.958 0.927 0.878 0.803
-50 0.9999 0.9997 0.999 0.997 0.994 0.988 0.977 0.958 0.927 0.878 0.803
-40 1.0000 0.9998 0.999 0.998 0.994 0.988 0.977 0.958 0.927 0.878 0.803
-30 1.0000 0.999 0.998 0.994 0.988 0.977 0.958 0.928 0.879 0.804
-20 1.000 0.998 0.995 0.989 0.978 0.959 0.928 0.879 0.804
-10 1.000 0.997 0.990 0.979 0.961 0.930 0.881 0.806
0 1.000 0.994 0.983 0.964 0.933 0.884 0.809
10 1.000 0.989 0.970 0.939 0.890 0.815
20 1.000 0.981 0.950 0.901 0.826
30 1.000 0.969 0.920 0.845
40 1.000 0.951 0.876
50 1.000 0.925
60 1.000
Table 3 Dilution coefficients
As an example: A gas sample is drawn from a 40 °C (Td) environment and introduced into a 10 °C (Td) enviroment, where
the measured gas concentration is 5.32%. In the 40 °C (Td) environment this corresponds to 5% CO2 (5.32% X 0.939= 5.00%)
since the higher water content diluted the sample.
5. Why is the CO2 concentration
reading higher than expected when
using the pump sampling method with
drying tubing?
When using the Nafion® tubing to dry the sample, the CO2
concentration of the dry sample will be slightly higher
than in the wet sample. This is due to a phenomenon
called dilution. The CO2 density is “diluted” in the
incubator by the volume that the water vapor occupies.
If water vapor is removed from the sample, the fractions
occupied by other gases, including CO2, will increase
accordingly.
Table 3 contains the dilution co-efficients for the gas
concentration when drying a gas sample. Dew point (at
1013 hPa) of the gas sample in the incubator is chosen on
the horizontal axis, and the dew point of the gas sample
at the measurement point is chosen on the vertical axis.
Dew point of the gas sample at the measurement point
can be determined with a humidity probe (HMP75B,
HMP76B or HMP77B).
|
|