You may also be interested in
Analyzer, Mass spectrometer, Gas analyzer, Hydrogen analyzer, Spectrometer
Text version of the page
2. 1. 3. Features Functions Benefits ? Fast and reliable Measurement? High Flexibility, Dynamic Range? Proven Sample Gas Inlet, Integrated sample conditioning System
1. Ion Source2. Sector Mass filter3. Particle Detector Principle of Operation Targeting the specific needs within the applications,the series H-SENSE is based on the Electron Pulse Ionization (EIMS) Mass Spec principle optimized to measure H 2 and He fast and reliably. By using the electron ion source the gas sampleions are energized, focused and separated fur- ther on in a magnetic field in order to detect only Hydrogen and Helium. The unique and fully automated gas inlet sys-tem compensates for pressure variations in between 500 mbar and 3 bar. The system guar- antees the correct measurement and avoids any discrimination of the gas.Any contamination due to condensation or par-ticulate matters is minimized by the integrated 2 micron fine filter in combination with a cross flow sample conditioning system, which takes care of any potential water content. Technical Data Mass range2, 3, 4 amu Lower detection limit0 .... 30.000 ppm and 0,1 .... 100 Vol.% Analisys time 4 Hz Measurement range Tel.: (734) 424-1574 H-Sense Hydrogen Gas Analyzer - 58717 ·Fax: (734) 424-1580e-mail: vfi@vandfna.com 1136 Baker RoadDexter, Michigan 48130·http://www.vandfna.com DB–H–SENSE / E / 04.03 Technical changes reservedmanufactured by |
|