1. Rlmary Ion Source 3. Charge Excange Cell 6. Vacuum System
2. Octopol Seperatlon Device 4. Quadrupol - Mass Filter 7. Gas Inlet System
5. Rartice Detector
Principle of Operation
Targeting the specific needs within the application, the model CO2-SENSE is based on the measurement principle called Ion Molecule Reaction (IMR).
By using the IMR technique, the measurement becomes flexible and easily adaptable to new applications, selec­tively and free of any interference. Unlike others, no frag­mentation or overlapping spectra can damage the inter­pretation of the detected results.
IMR means using primary ions with lower energy level (10 eV .. 14 eV) to completely ionize the probe gas mol­ecules. The signal/noise ratio will be optimized by the integrated Octopole separator focusing the primary ions and filtering out any interferences.
The quadrupole mass filter (0-500 amu) then separates the molecules for further detection at the fast pulse counter. The temperature and pressure compensated sample gas inlet guarantees the correct measurement and avoids dis­crimination of the gas. Any contamination due to con­densation or particulate matters is minimized by the unique sample inlet design.
Features Functions Benefits
• Multi component gas analyzer system, Lowest detectable limits,
• Robe sampling via standard pressureless PTFE tubing, Fast response time, • Flexible operation, User friendly adaptation for additional gases to measure
Specification CO2 Quality Control
Components
Specification
EGA in ppm v/v
LDL CO2-SENSE in ppm v/v
Components
Specification
EGA in ppm V/V
LDL
CO2-SENSE in ppm v/v
carbon dioxide oxygen ammonia carbon monoxide methane benzene toluene xylene nitric oxide nitrogen dioxide sulfur dioxide hydrogen sulfide carbon oxysulfide
Technical Data
99,9 %
< 30
< 2,5
< 10
< 50
< 0,02
4
0,25
5
1
1 ppb
1 ppb
2 ppb
0,09 0,04 0,1
0,06
0,02
carbon disulfide methanethiol ethanethiol total sulfur
water
halogenated HC's chlorinated HC's phenol
hydrogen cyanide
vinylchloride
acetaldehyde
methanol
ethanol
phosphin
< 0,1
< 50
0,02 0,02 0,02
< 0,1
10
0,17
0,05
0,05
0,1
0,1
0,1
1
1
0,1
< 2,5
< 2,5
< 1,0
< 0,1
< 0,1
< 0,2
< 10
< 0,3
Mass range Analysis time Measurement range Response time Lower detectable limit Drift
Ambient temperature Temperature change
0 - 150 amu
10 - 6500 msec/amu
104
T90 < 50 msec
< 1 ppb (Benzene in air)
< ± 5% over 12 h (1 ppm Benzene) 20°C - 40°C,
max 1°C/ h
Humidity
Gas Inlet temperature Multi sampling Gas consumption Rower Dimensions Weight
max. 80 % (none condensng) 80°C - 190°C
Optional 12x Multi sampling
30 - 250 ml / min
220V / 50Hz (115V / 60Hz), 800 W
590 x 650 x 730 mm
87 kg