Sarasota Liquid Density Meters - Thermo Scientific - Process Systems - #4

/ 4


catalogue search
P. 01
P. 02
P. 03
P. 04
Pages:


See other catalogues for Thermo Scientific - Process Systems

Text version of the page

Specifications

Sarasota Liquid Density Meters - 23500 Thermo Scientific Sarasota Liquid Density Meters

Functional Specifications Transducer Calibration AccuracyAvailable to ±0.1 kg/m
3 (±0.0062 lb/ft
3 ) Repeatability0.02 kg/m
3 (0.0012 lb/ft
3 )Flow RangeVertical installation: 0 l/min to 300 l/min (0 USG/min to 79 USG/min); Horizontal installation: 5 l/min to 300 l/min (1.3 USG/min to 79 USG/min) Operating Density Range0 kg/m
3 to 2100 kg/m
3 (0 lb/ft
3 to 131.1 lb/ft
3 ) InstallationVertical installation (standard), horizontal installation (optional); No instrument or pipework supports required Pressure Effect (corrected)0.003 kg/m
3 /bar (0.000013 lb/ft
3 /psi) note: correction coefficients applied Temperature Effect (corrected)0.005 kg/m
3 /°C (0.0002 lb/ft
3 /°F) note: correction coefficients applied Density Meter DimensionsSee dimensional diagrams Shipping Dimensions590 mm x 390 mm x 290 mm (approx 24 in x 16 in x 12 in) Net Weight11 kg (24 lb) Shipping Weight15 kg (33 lb) Environmental RatingIP65 (NEMA 4X) Electrical ConnectionsScrew terminals; Cable entry: 2 x
3 ⁄
4 -in NPT Temperature MeasurementHigh accuracy
1 ⁄
3 DIN integral 4-wire PT100 Local Display (H option)4
1 ⁄
2 -digit 7.6 mm (0.3 in) 7-segment LCD display. Resolution 0.1% or 0.01% depending on display variable. Secondary ContainmentAs flange rating to Class 300 then 2.5 times maximum safety flange rating to Class 600 Factory Calibration Range650 kg/m
3 to 1600 kg/m
3 (40.58 lb/ft
3 to 99.98 lb/ft
3 ) Ambient Temperature Range-20°C to +60°C (-4°F to +140°F) ambient Process Temperature RangeSarasota FD910 / FD950 / FD960: -50°C to +180°C (-58°F to +356°F)Output F option (frequency output): Frequency related to density on 2-wire current modulated loop 6 mA to 18 mA; 4-wire PT100 Hoption (headmounted electronics): Analog 4-20 mA related to density or density derived variable; HART protocolPowerSupplyF option (frequency output): 13-28 VDC 10 mA average (peak 18 mA) Hoption (headmounted electronics): 2 x 13-28 VDC 25 mA; 4-20 mA current pressure input available Maximum Operating PressureSarasota FD910 / FD950 / FD960:as flange rating Material Specifications SensorSarasota FD910 stainless steel (316L / 1.4404); Sarasota FD950: Hastelloy C276; Sarasota FD960: Ni-Span C Other Wetted PartsSarasota FD910 (Class 150, 300) / FD960:316L stainless steel (316L / 1.4404) Sarasota FD910 (Class 600) / FD950: Hastelloy C276 CaseStainless steel (316L / 1.4404) Electronics HousingCopper free aluminum grey epoxy finish; Plate glass window for local display option Process Connections 1-in ASME B16.5 RF (raised face)Sarasota FD910 / FD960: stainless steel (316L / 1.4404) – Class 150, 300 or 600Sarasota FD910 / FD960: duplex (A 182 Gr.F51) – Class 150, 300 or 600 Sarasota FD950 only: Hastelloy C276 – Class 150, 300 or 600 25-mm BSEN1092 RF (raised face - type B)Sarasota FD910 / FD950 / FD960: Up to maximum PN100 Other Flange TypesConsult Thermo Fisher Compliance/Certification Quality AssuranceISO 9001:2000 CE markCompliant Electromagnetic Compatibility Compliant (EN 61326:1997) Pressure Equipment Directive (97/23/EC)Sarasota FD910 / FD950 / FD960: category III Low Voltage DirectiveCompliant Safe Area UseAs standardBS EN ISO 15156 / NACE MR0175 ConformanceAvailable on Sarasota FD910 and Sarasota FD950 onlyATEX Conformance: Intrinsically Safe F option (frequency output): Ex II 1 G EEx ia IIC T6 (-20°C
<_ Ta



©200 8 Thermo Fisher Scientific Inc. All rights reserved. HART is a registered trademark of the HART Communication Foundation. Tri-Clamp is a registered trademark of Tri-Clover, Inc. Hastelloy is a registered trademark of Haynes International, Inc. Ni-Span C is a registered trademark of the Special Metals Corporation.All other trademarks are the property of Thermo Fisher Scientific Inc.and its subsidiaries. Results may vary under different operating conditions. Specifications, terms and pricing are subject to change. Not all products are available in all countries. Please consult your local sales representatives for details. Literature Code PI.2024. 11 08 Process Instruments www.thermo.comRoom 1010 - 1019 Ping’an Mansion No. 23 Jinrong Street Xicheng Dist, Beijing 100032 CHINA A-101, ICCTrade Tower, Senapati Bapat RoadPune 411016 Maharashtra, INDIAIon Path, Road Three, WinsfordCheshire CW7 3GA UNITEDKINGDOM1410 Gillingham LaneSugar Land, TX 77478 USA +86 (10) 5850-3588+86 (10) 6621-0847 fax+91 (20) 6626 7000+91 (20) 6626 7001 fax+44 (0) 1606 548700+44 (0) 1606 548711 fax+1 (800) 437-7979+1 (713) 272-0404 +1 (713) 272-4573 fax

pageCatalog pdf di En 2012-06-22-01