1. Catalogs
  2. Thermal Detection
  3. Validation Thermocouple Looms

Validation Thermocouple Looms

Validation Thermocouple Looms

Product catalog summary
Overview: The document discusses the use of Copper/Constantan type ‘T’ thermocouples for temperature measurement in pharmaceutical validation, specifically within looms for chamber mapping.
Loom Description: A loom facilitates the handling of multiple long thermocouples. It is custom-made based on customer specifications and is designed for easy transportation and installation.
Wire Types: The document outlines three main types of wire insulation used in thermocouples:
  • PFA Extruded: Smooth and easy to clean, with optional weep holes to prevent condensate issues.
  • PTFE Lapped: Features tightly wrapped insulation to minimize capillary action.
  • Kapton®: Suitable for high-temperature applications, with optional stainless steel tips for added protection.
Optional Features: Various optional extras are available, including:
  • Encapsulated hot junctions to reduce oxidation and condensate penetration.
  • Identification at both ends of the thermocouple.
  • Stainless steel tips for Kapton® wires.
  • Mini plugs for connectivity.
Specifications: The document provides a detailed table of specifications for different wire types, including:
  • Insulation material and conductor format.
  • Maximum temperature ratings (250°C for PFA and PTFE, 300°C for Kapton®, with short-term capability up to 350°C).
  • Availability of features like bare or encapsulated hot junctions, heat shrink, and weep holes.
Contact Information: The document concludes with contact details for Thermal Detection Limited, including address, phone, fax, email, and website.
See more

Catalog excerpts

Validation Thermocouple Looms-1

Validation Thermocouples in Looms Product Code: VTFP-LOOM Description: The preferred method of temperature measurement for pharmaceutical validation purposes is the Copper/Constantan type T thermocouple. A loom allows for easy placement and handling of a large number of long thermocouples when mapping a chamber. The loom is produced by bundling and securing pre-cut thermocouples, number and length as specified by the customer. The loom is reeled for ease of transportation to the customers site. Once the loom and hot junctions are in place, the cable ties can be removed by the engineer. The main wire types are: • PFA Extruded : Smooth, easy to clean. • PTFE Lapped: Tightly wrapped insulation minimises capillary action. • Kapton®: For high temperature applications such as depyrogenation. Weep holes can be specified for PFA insulated wires to counter any condensate travelling up the inner jacket of the thermocouple by capillary action. Encapsulation of the hot junction also reduces the incident of condensate penetration and further ensures that the rate of oxidisation of the thermocouple hot junction is reduced. For Kapton® wire a thin wall stainless steel tip can be specified for added hot junction protection. Available Optional Extras Conductor Max Temp Bare Hot Format °C Junction Teflon Encapsulated Hot Junction * Kapton® can go to 350°C for short periods Datasheet : 021 Thermal Detection Limited Unit 6, Primrose Hill Industrial Estate Orde Wingate Way Stockton-on-Tees, England, TS19 0GA Email: [email protected] Web: www.thermal-detection.com

 Open the catalog to page 1

All Thermal Detection catalogs and technical brochures

  1. PT100

    2  Pages

  2. MD series

    3  Pages

  3. LD Series

    3  Pages

*Prices are pre-tax. They exclude delivery charges and customs duties and do not include additional charges for installation or activation options. Prices are indicative only and may vary by country, with changes to the cost of raw materials and exchange rates.