DATA SHEET No. 002.0
PRODUCT DESCRIPTION Autoclave manufacturer Chamber/Load Probe PRODUCT CODE SL100
The environmental conditions found in pharmaceutical autoclave chambers are harsh and extremely testing for any equipment. These same conditions are a major cause of sensor failure in chamber/load probes. This is frequently due to the ingress of moisture at the detector and the unacceptable emission of steam/condensate through the bulkhead connection. Additionally in water cycle systems the chamber assembly is subjected to a continual cascade of high purity water at an elevated temperature and pressure. The ‘Steriprobe’ has been designed to withstand these conditions and offers a reliability factor surpassing that previously obtainable. During manufacture every weld undergoes NDT examination and each completed assembly is 100% tested.
S P E C I F I C A T I O N
MECHANICAL Sheath dimensions : 3.0mm, 4.7mm or 6.0mm OD x 100mm long with 50mm increments up to 300mm long and in 316 stainless steel. Flexible conduit : 10.1mm OD x 1500mm long and in 316 stainless steel. Bulkhead tail piece : 6mm OD x 150mm long and in 316 stainless steel. >
ELECTRICAL Resistance Thermometer Detector : single, duplex or triplex PT100 with 3 or 4 wire connection, Class ‘A’ or fractional DIN tolerance and to BS EN BS EN 60751:1996 Flying lead : PTFE insulated copper conductors, 7/0.16mm (26 AWG) to ASTM-B-298 with 250V RMS rating and with a heavy duty silicone rubber jacket and an overall length of 3000mm. Operating range : -50 to +150 ° C.
Thermocouple Element : single, duplex or triplex elements, types ‘T’ (Cu/ Con), ‘E’ (NiCr/Con), ‘J’ (Fe/Con) or ‘K’ (NiCr/ NiAl) in class 1 material and to BS EN 60584-1:1996 . Lead wire : PTFE insulated conductors, 1/0.315mm (28 AWG) and 7/0.16mm (26 AWG) and with a heavy duty silicone rubber jacket, overall length of 3000mm. Operating range : - 50 to +150 ° C. >
®