SD050 Drain/Air Detector Probe - 1

Autoclave Drain and Air Detector / Vent Probe Description: Although the application conditions for the drain or air detector are not as demanding as those presented to the autoclave chamber/load probe, their integrity is just as critical. To maintain this integrity both types are subjected to the same manufacturing programme as are the chamber/load probes. Specification: SHEATH DIMENSIONS Diameter : Metric: 3.0mm, 4.0mm, 4.5mm, 5.0mm, 6.0mm, 8.0mm Imperial: ⅛”, 3/16“, ¼” Length: 100mm to as required 316/316L stainless steel RTD OPTION Single, duplex or triplex PT100 with 2, 3 or 4 wire connection, Class ‘A’ or fractional DIN tolerance and to BS EN 60751:2008. PFA insulated copper conductors, 7/0.15mm to ASTM B298 with a 250V RMS rating as standard and contained within an PFA jacket with a minimum overall length of 1000mm. OPERATING RANGE Thermal Detection Limited Unit 6, Primrose Hill Industrial Estate Orde Wingate Way Stockton-on-Tees, England, TS19 0GA PFA insulated type ‘T’ conductors, 1/0.3mm or 7/0.2mm as standard in Class 1 material. Single thermocouple construction, flat pair with a PFA insulation and jacket. Duplex thermocouple construction, twisted pairs with a PFA insulation and jacket. Triplex thermocouple construction, three individual flat pairs, each with PFA insulation and contained within a silicone sleeve. Minimum overall length of 1000mm. OPERATING RANGE Single, duplex or triplex element, type ‘T’ (Cu/Con) to BS EN 60584-1:2013. Type ‘K’ (NiCr/NiAl) also available PROCESS CONNECTIONS See our range of compression fittings (MSC). Sheath can be supplied with a welded on Tri-clamp flange also in 316L stainless steel. Fax : +44 (0) 1642 618307 Tel : +44 (0) 1642 602878 Email: sales@thermal-detection.com Web: www.thermal-detection.com © T

Open the catalog to page 1
SD050 Drain/Air Detector Probe - 2

SD050 Autoclave Drain / Air Detector /Vent Probe Product Enquiry Form Please download PDF and fill in the boxes. Return form to sales@thermal-detection.com or fax to +44 1642 618307 Organisation: Country Section 1 - unit dimensions (in mm) Section 2 - Sensor Configuration Sensor Number Thermocouple Only Hot Junction Single Duplex Triplex Section 3 - Optional Features (in mm) Reduced Probe Tip Section 4 - Application Details Number of probes required Additional information Make of autoclave Industry Type Thermal Detection Limited Unit 6, Orde Wingate Way Primrose Hill Industrial Estate...

Open the catalog to page 2

All Thermal Detection catalogs and technical brochures

  1. PT100

    2 Pages

  2. MD series

    3 Pages

  3. LD Series

    3 Pages