You may also be interested in
Communication module, Interface, Recorder, Syringe pump, Spectrophotometer
Text version of the page
Foxy R2
68-2130-006
High Flow Foxy R2
68-2130-007
Foxy R1
68-3870-010
Retriever 500
68-3880-001
Inlet Tubing
and Flow Rates
1/8" OD (0.062" ID),
25 125 mL/min
(up to 25 mL/min
optional)
1/4" OD (0.125" ID),
up to 1 L/min
1/16 OD (0.031" ID),
up to 25 mL/min
(25 125 mL/min
optional)
1/16" OD (0.031" ID),
up to 25 mL/min
(25 125 mL/min optional)
Maximum Tube Height 196 mm 180 mm 160 mm 180 mm
Racks 2, with RFID 2, with RFID 1, no detection 17, with infrared rack sensor
Maximum Tube Qty 288 12/13 mm tubes 144 12/13 mm tubes 102 12/13 mm tubes
Max Fraction Volume Unlimited with funnel racks 50 mL with 29 mm centrifuge tubes
Construction Materials:
Racks
Funnel Racks
Diverter Valve
Construction
Drip Tray
Drop Former
Stainless steel and polypropylene
ECTFE (Halar) or ETFE (Tefzel) funnels
PEEK and FFKM (fluoroelastomer)
Cast Aluminum
Polypropylene
PEEK
Polypropylene
n/a
PCTFE (Kel-F*)
Polypropylene
Polypropylene and stainless steel
ECTFE (Halar) or ETFE (Tefzel)
Solvent Compatibility Aqueous and Organic Aqueous (Organic with optional valve)
Test Tube Time Change 0.10 0.58 seconds
(dependent on collection vessel tube advance speed)
0.35 seconds with 12/13 mm rack, within
rack; 0.57 seconds between racks
Memory Eight methods Current program
Collection Modes:
Time
00:00:01 99:59:59, 1 second intervals 5 999 sec; 0.1 99.9 min. in 0.1 min units;
1 999 min in 1 min units
Drop 1 999 drops per fraction 1 999 drops per fraction;
up to 7 per second
Volume Up to 999 units Up to 999 units w/Tris Pump
Peak Detection Slope, Threshold, four time windows,
Combination of any of the above collection modes
External
System Control Front panel touch screen, Ethernet or RS-232 Front panel touchkey
Inputs Pump counter, external peak, inject, tube advance: +3.3 to 5 VDC or contact
closure, active low, 30 ms minimum, 5 VDC max. Analog peak voltage: 0 to
1.0 V nominal, 1.2 to + 1.2 V usuable, 5 to +5 V max.
18 volts, Max. Pump counter: CMOS or
TTL logic 0 for 7.7 ms min, Max rate 60
counts per sec. Tube change: Open collector
TTL logic 0 or contact closure. Pause:
Selectable; inactive, active TTL logic 1, or
active TTL logic 0
Outputs Pump pause/stop and event (tube change), 300 ms:
open collector, active low, 15 VDC/100 mA maximum.
Event marks: Open collector active for tube
or rack change
Automatic Shut-off Rack full, last tube, end of time window or external signal Last tube
Security Valve (2-way) n/a Optional
Diverter Valve (3-way) Included Optional
Valve Pressure Rating 0 28 psi 0 28 psi
Temperature Range 0 40 ΊC, 32 104 ΊF 0 40 ΊC, 32 104 ΊF
Cold Room Compatible Yes Yes
Dimension: WxHxD 12.25 x 14.9 x 21.0 in
31.1 x 37.8 x 53.3 cm
12.25 x 15.5 x 21.0 in
31.1 x 39.4 x 53.3 cm
12.25 x 14.0 x 13.0 in
31.1 x 35.5 x 33.0 cm
8.3 x 7.09 x 11.4 in
21.0 x 18.0 x 29.0 cm
Weight 10.3 kg (22.8 lbs) 10.4 kg (22.9 lbs) 7.1 kg (15.6 lbs) 3.0 kg (6.61 lbs)
Power Requirements 100 240 VAC, 50-60 Hz, 1 ampere max.
Includes one IEC power cord supplied for regional AC power
(100, 117, 234 VAC)
14 VAC, 50/60 Hz, 150 mA max. from
Teledyne Isco 15W wall transformer for
regional AC power (100, 117, 234 VAC)
Certification CE, CSA CE
Fraction Collector Selection Guide
* A registered trademark of 3M
To order contact Teledyne Isco or your local representative
4700 Superior Street Lincoln, NE 68504 USA
Toll free: (800) 228-4373 (USA & Canada) Phone: (402) 464-0231
Fax: (402) 465-3022 iscoinfo@teledyne.com
Brochure L-5146 Printed in the USA © 2009, all rights reserved 10/09
|
|