SECTION 4 Calibration

4.1 Hastings 760 Plus Piezo Transducer manufacturer Zero and Span Adjustment

1.Install the transducer and power supply/display then apply operating power. Wait a minimum of 15 minutes before proceeding.2.Evacuate system to a pressure less than 0.010 Torr (1.00E-02 Torr). 3.Adjust transducer zero trimpot (Fig. 2-1) until the power supply/displayindicates 000.0 ±000.1 Torr.4.Increase system pressure to atmosphere and adjust span trimpot until the power supply/display indicates the known local atmospheric pressure ±000.1 Torr.5.Repeat steps 2 through 4 if necessary.

4.2 Hastings 760s Option-5 Transducer Zero and Span Adjustment

1.Install the transducer and apply operating power. Wait a minimum of 15 minutes beforeproceeding.2.Evacuate the system to a pressure less than 0.01 Torr (1.00E-02 Torr). 3.Adjust transducer zero trimpot (Fig. 2-1) until display indicates 000 ± 001 Torr. 4.Increase system pressure to atmosphere and adjust span trimpot (Fig. 2-1) until displayindicates the known local atmospheric pressure ±001 Torr.5.Repeat steps 2 through 4 if necessary. page 13
DirectIndustry's Virtual Technical Library: PDF Catalogue | Technical Documentation | Brochure | Manual | Industrial directory | Specifications | Characteristics
Search Go
page 1 p.1
page 2 p.2
page 3 p.3
page 4 p.4
page 5 p.5
page 6 p.6
page 7 p.7
page 8 p.8
page 9 p.9
page 10 p.10
page 11 p.11
page 12 p.12
page 13 p.13
page 14 p.14
page 15 p.15
page 16 p.16
page 17 p.17
page 18 p.18
page 19 p.19
page 20 p.20
page 21 p.21
page 22 p.22
pdf-page pdf di En 2010-03-11-15