3010TA - 1

TELEDYNE ANALYTICAL INSTRUMENTS ADVANTAGES • • • • • Linearity of analysis across three user-programmable ranges AutoRanging capabilities Digital interface allows monitoring from a remote station Extended-life, maintenance-free sensor Comprehensive self testing function STANDARD FEATURES MODEL 3010TA Process Trace Oxygen Analyzer eledyne’s Model 3010TA Trace Oxygen Analyzer is a versatile microprocessor-based instrument for detecting oxygen at the parts-per-million (ppm) level in a variety of gases. The 3010TA is designed as a “split architecture” instrument, meaning that a general purpose Control Unit for non-hazardous areas controls a specially designed Analysis Unit or “remote probe” that can operate in hazardous areas. Simple menu choices, membrane command switches and dual LCD and LED displays make set-up and operation clear and quick. Three User-Configurable Analysis Ranges Three user-configurable ranges are standard, with an excellent linearity precluding the need to recalibrate when changing ranges. Two fully programmable concentration alarms provide the versatility to satisfy nearly any requirement. All features offer a sophistication that assures the 3010TA will provide years of service. Convenient Outputs For Data Two standard 0-1 VDC outputs and two 4-20 mADC isolated outputs provide both concentration and range identification. A bi-directional RS-232 serial interface is incorporated to relay information to a host computer for remote monitoring of critical functions. • An explosion proof NEMA 4/7 rated analysis unit enclosure • A 2-line alphanumeric display screen, driven by microprocessor electronics continuously prompting and informing the operator • High resolution, accurate readings of oxygen content from low ppm levels through 25% • Stainless steel cell block • Advanced Micro-fuel Cell designed for trace analysis • Versatile analysis over a wide range of applications • Microprocessor based electronics: 8-bit CMOS microprocessor with 32kB RAM and 128kB ROM • Three user-definable output ranges (from 0-10 ppm through 0250,000 ppm) assuring a perfect match for the user’s process and equipment • Air-calibration range for convenient spanning at 20.9% • Auto-Ranging automatically selects the proper preset range for a given measurement. Manual override allows the user to lock onto a specific range of interest. • Two adjustable concentration alarms and a system failure alarm • Extensive self-diagnostic testing at start-up and on demand with continuous power-supply monitoring • Two way RS-232 serial digital port for use with a computer or other digital communication device • Four analog outputs; two for measurement (0-1 VDC and Isolated 420 mADC) and two for range identification Monitoring inert gas blanketing Air separation and liquefaction Chemical reaction monitoring Semiconductor manufacturing Petrochemical process control Quality assurance Gas analysis certification Built for Reliability and Performance

Open the catalog to page 1
3010TA - 2

MODEL 3010TA TRACE OXYGEN ANALYZER SPECIFICATIONS: Ranges: 3 customer programmable ranges (minimum 0-10 ppm) with AutoRanging Calibration range: 90% of FS at 77°F (25°C) in less than 65 seconds Operating temperature: Signal output: Analytical measurement: 0-1 VDC and 4-20 mADC (isolated) Sample connections: 1/4” with conversion to 6mm available Area classifications: Analysis unit: Explosion proof enclosure is U/L and CSA listed for Class I, Division 1, Groups B, C, D service NEMA 4/7 rated Control unit: General purpose flush panel mounted Integrally mounted cal/zero valves Plumbed for...

Open the catalog to page 2

All Teledyne Analytical Instruments catalogs and technical brochures

  1. PEM 9004

    4 Pages

  2. INSTA TRACE

    2 Pages

  3. PHOTO-X

    6 Pages

  4. MODEL MAX5

    2 Pages

  5. 9110T Series

    4 Pages

  6. 9060Z

    4 Pages

  7. 8800 series

    6 Pages

  8. 1220

    4 Pages

  9. SERIES 6700

    6 Pages

  10. model 9070

    2 Pages

  11. MODEL 3010TA

    2 Pages

  12. 3000TA-XL

    2 Pages

  13. BDS series

    4 Pages

  14. 3020T

    2 Pages

  15. 3000TB

    2 Pages

  16. R-15

    1 Pages

  17. UFO-130-2

    2 Pages

  18. T-7

    1 Pages

  19. R-22MED

    1 Pages

  20. R-17MED

    1 Pages

  21. PEM-9004

    4 Pages

Archived catalogs