LCR 200 SPECIFICATIONS
DC Voltage Measurement:
Range
2V. 20V. 200V and 600V
Resolution
ImV. lOmV. lOOmV. IV
Accuracy
± (0.4%+3digit)
Input Impedance
IMQ
AC Voltage Measurement (True RMS):
Range
2V. 20V. 200V and 600V
Resolution
lmV.IOmV.IOOmV.lv
Accuracy
± (0.8<7r+5digit) @ 40 Hz -
1 KHz
Input Impedance
1 MQ
DC Current Measurement:
Range
2A. 0.2A. 20mA and 2mA
Resolution
1mA, 0.1mA. 0.01 mA and 1
uA
Accuracy
±(0.4%+3digil)
Current Shunt
0.IQ<» current > 20mA I0Q @ current < 20mA
AC Current Measurement (True RMS):
Range
2A. 0.2A. 20mA and 2mA
Resolution
1mA, 0.1mA, 0.01 mA and 1
UA
Accuracy
± (0.8%+5digil) @ 40 Hz -
1 KHz
Current Shunt
0.1 Q @ current > 20mA 10Q@ currents 20mA
Continuity Check:
Continuity Threshold
50Q (Approximately)
Continuity Indicator
Tone Buzzer
Input Protection
600V (max)
Diode Test:
Test current
2.5mA (Typical)
Open circuit voltage
5.0VDC (nominal)
Input protection
600V (max)
General:
Temperature
0 to 40° C (Operating) -20 to 70° C (Storage)
Relative Humidity
Up to 85%
AC Power
ll()V/220V± 10%, 60/50H2
Dimensions
300mm x 220mm x 150mm
( H.8"x8.7"x5.9")(DxWxH)
Weight
4.5 kgs
Standard Accessories:
©TL09C Kelvin clips
pe-
© Test leads
pe
Test condition:
Frequency : IOOHz/1201-iz/l KHz/
l()KHz/l()0KHz/200KHz
Level : I Vrms/250mVrnis/
50mVrms/l Vdc (DCR only)
Measuremeni Parameters : Z. Ls. Lp. Cs. Cp. DCR.
ESR. D. Q and 9
Basie Aecuraey : (). I
Dual Liquid Crystal Display with LED baeklil
Auto Range or Range Hold selection
Open/Short Calibration
RS232C Interface
Primary Parameters Display:
Z : AC Impedance
DCR : DC Resistance
Ls : Serial Inductance
Lp : Parallel Inductance
Cs : Serial Capacitance
Cp : Parallel Capacitance
Secondary Parameters Display:
: Angle Impedance
ESR : Equivalent Series Resistance
D : Dissipation Factor
Q : Quality Factor
Combinations of Display:
Serial Mode : Z-9. Cs-D. Cs-Q, Cs-ESR,
Ls-D. Ls-Q. Ls-ESR Parallel Mode : Cp-D. Cp-Q. Lp-D. Lp-Q
DMM Functionalities: DCV. ACV. DC A. ACA. Diode/Continuity Check Measurement Range:
Parameter
Range
Z
0.000 £2 to 500.0 MQ
L
0.030 //H to 9999 H
C
0.003 pF to 80.00 mF
DCR
0.000 Q to 500.0 MQ
ESR
0.000 Q to 9999 Q
D
0.000 to 9999
Q
0.000 to 9999
9
-180.0° to 180.0°
V
O.OOOmV to ± 600.0V
I
0.000mA to ± 2.000A
Optional Accessories:
® TL09A SMD test probe ©TL09B Component fixture
Specifications subject to change without notice.

pageCatalog pdf di En 2012-06-22-01