Techne Industrial Fluidized bath basic system layout
Stack or existing exhaust ventilation
Techne does not supply interconnecting pipework which is typically 4" diameter.
7030772 Cast aluminum Extraction fan 350 CFM, 240 volt single or 3 phase
Air damper (not supplied by Techne)
Parts basket would be lowered into the bath below the lid.
6036156 6036157 _Lid
Extraction collar
IFB
Fluidized Bath
7031154 Cyclone
IT
TT
With parts in the bath and lid on, the damper would be adjusted so fumes and smoke are pulled away from the bath and into the exhaust system. If the damper is open too much it would create a negative pressure in the bath working area resulting in significant heat and sand loss. The purpose of the Cyclone is to recover any sand pulled out and heavier deposits which can be sieved and reused.