PRESSURE CONNECTION
MS33649-4
WITH %" HEX
-4V4"-
#0082 MOUNTING BRACKET (OPTIONAL)
SPECIFICATIONS
0-200, 0-250, 0-300, 0-500, 0-750, 0-1000, 0-1500, Thermal Zero Shift 0-2000, 0-3000 PSIA or PSIS (14.7 PSIA reference). _
All fluids compatible with Carpenter 20 and 347 Triaxial
stainless steel._ Mechanical Shock
H-*?0,15. £¿£1 Citper vo,t excitation- Acceleration Error Calibrated at 10.00 VDC excitation.__ -
Less than ±0.010% FSO per °F over compensated temperature range.
30 ffs applied for 11 milliseconds will not cause change in performance characteristics.
Along most sensitive axis: ±0.005% FSO per G._
10 volts DC or AC recommended. 15 volts DC or AC maximum._
350 ±3.5 ohms resistive at +70°F. Input circuitry symmetrical.
350 ±5.0 ohms.
Greater than 10,000 megohms at 50 VDC between all terminals in parallel and case at +70°F.
7/16-20 internal thread per MS33649-4 is standard. Options available._
0-200 PSI range...................0.04 cu. in.
0-250 PSI range...................0.04 cu. in.
0-300 PSI range...................0.04 cu. in.
0-500 PSI range...................0.04 cu. in.
0-750 PSI range...................0.04 cu. in.
0-1000 PSI range...................0.04 cu. in.
0-1500 PSI range....................0.04 cu. in.
0-2000 PSI range...................0.04 cu. in.
0-3000 PSI range...................0.04 cu. in.
Hermetic Bendix Pygmy Types PCIH-10-6P is standard. Standard wiring: Excitation +D, -C & F; Signal +A, -B & E.
Options available. _
Entire housing and pressure cavity of stainless steel. All electrical components sealed against adverse environmental conditions.
Less than 9 ounces.
J"ressure Ranges
Measurand Fluids
Full Scale Output
Excitation
Zero Balance
0.00±0.03 MV per volt at +70°F._
Within ±0.25% FSO._
End Point Linearity
Input Resistance
Less than 0.25% FSO.
Hysteresis
Repeatability
Within 0.10% FSO.__
Output Resistance
Resolution
Infinite.__
Insulation Resistance
0-200 PSI range........... 4.0KHz _
0-250 PSI range........... 4.5KHz
0-300 PSI range........... 4.8KHz Pressure Connection
0-500 PSI range........... 6.3KHz
0-750 PSI range........... 7.2KHz
Natural Frequency
Pressure Cavity Volume
0-1000 PSI range........... 7.5KHz
0-1500 PSI range...........10.5KHZ
0-2000 PSI range...........U.3KHz
0-3000 PSI range...........12.0KHz
Application of 3 times rated pressure range or 6000 PSI, whichever is less, will not cause change in performance characteristics.
Greater than 5 times rated pressure range or 8000 PSI
whichever is less. Electrical Receptacle
-30°Fto +170°F. Options available.
Proof Pressure
Burst Pressure Ratini
Compensated Temperature Range
Operating Temperature Range
Enclosure
-100-F to +300°F.
Less than ±0.005% FSO per *F over compensated -
temperature range. Weight
Thermal Sensitivity Shift
Terminology in accordance with ISA Standard S37.3.