| | | Synkera Technologies, Inc. 2021 Miller Dr., Suite B Longmont, CO 80501 | | |
| | | Low Level NOx Data Sheet (P/N 706) | | |
| | | |
| | | |
| | | SENSOR FEATURES: • Sensor detects very low NOx concentrations (<0.5 ppm to 10 ppm) • Larger dynamic range available (5 ppm to 100 ppm, may require larger power input to heater) • Environmental temperature range of — 20 to 50°C • Environmental humidity range of 0 to 90% RH, non-condensing • Low dependence on flow rate | | |
| | | Typical NOx Sensor Selectivity | | |
| | | |
| | | | | | | | | | | Challenge | Concentration | Rg/Ra or | | | | Gas | | | | | | NO2 | 10 ppm | 200 | | | | NOx | 10 ppm | 45 | | | | H2 | 100 ppm | -2.2 | | | | NH3 | 25 ppm | -1.1 | | | | CO2 | 5% | No response | | | | SO2 | 5 ppm | No response | | | | CH4 | 1% | No response | | | | | | | | | | |
| | | |
| | | |
| | | Sensor Response Characteristics The figures below show typical response data for sensors operated in clean, dry gas. | | |
| | | |
| | | 1.0.E+09 | | |
| | | |
| | | 1.0.E+08 4 | | |
| | | S | | |
| | | |
| | | |
| | | ♦ NO2 ■ NOx Air | | |
| | | 1.0.E+07 | | |
| | | S in 'ill cd | | |
| | | |
| | | |
| | | 1.0.E+06 4 | | |
| | | |
| | | |
| | | 1.0.E+05 | | |
| | | |
| | | 0.1 1 10 100 Gas Concentration (ppm) | | |
| | | |
| | | Sensor response to NOx. Challenge gas is applied at 300s and 900s. | | |
| | | Sensitivity of NOx sensor operated at ~250°C. | | |
| | | |
| | | |
| | | Electrical Characteristics The electrical properties below are typical for the Low Level NOx Sensor prototypes. If the actual values differ the customer will be notified with the shipment. Circuits are available that will be preset to the correct values. | | |
| | | |
| | | | | | | | | | | | Property | Symbol | Value | Remarks | | | | Heater Power Consumption | Phl | ~ 500 mW | At Vh = 4.7 | | | | Heater Voltage | Vhl | 4.7 VDC | Tsensor ~200°C | | | | Heater Resistance | Rh | 32Q ± 2 Q | At room temperature | | | | Sensing Voltage | Vc | 5.0 VDC | Recommended | | | | | | | | | | | |
| | | |
| | | 720-494-8401 e-mail: info@synkera.com | | www.synkera.com 720-494-8402 (fax | | |
| | | |
| | | |
| | | For information on warranty, please refer to Synkera Technologies, Inc. Standard Terms and Conditions. Information on this data sheet represents typical values from a number of Synkera sensors. Actual values from sensor to sensor can vary slightly. | | |
| | | |
| | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | |