instruments >
CIVI
Associated to analog to digital converter: 800x family Low-cost instant viscosity
ADVANTAGES >
measurement
PRINCIPLE Especially developed for industrial processcontrol at ambient temperature The active part of the sensor is avibrating rod, held in oscillation by a
constant electrical power. No wear part, Low maintenance Operates in any position Simple, rugged, accurate and RELIABLE Easy-to-handle electronics, equipped with
standardized outputs and adjustablecalibration. Easy connects to any dataacquisition. OPERATING CONDITIONS Process Temperature: up to 40 °C >
N Working temperature: 120 °C max >
S Working pressure: 20 bars Watertightness of the body: IP 67 Material: 316 L stainless steel The vibration amplitude varies accordingto the product’s viscosity where the rod
is immersed. The signal
industrial converter associated Approved ATEX EEx d IIB T4..T6 MEASUREMENT RANGE Any range from 0-10 mPa.s to 0-10 000 mPa.s >
(800xfamily)
Repeatability: 0,2 % FS (warranted for T 40 °C) allows the vibration andprocesses the amplitude variations inorder to deliver a calibrated viscosityresponse. Weight : 3.4 kg >
Quality System MOUNTING >
In line On circulation pot On vessel 60 mini 1/4" or 1/2" Mounting flange to be welded (provided) 123 65 162 131 79 Ref. : 136/6