| | Pt 100 Pt 500 Pt 1000 | 2, 3, 4 wire | -90.00 °C to 850.00 °C | +/-0.005 % FS +/- 0.01 °C | |
| | Connection possibility through 4 mm connections free of thermal voltage | |
| | | |
| | Type K | NiCr-NiAl | -90.00 °C to 999.99 °C 1000.0 °C to 1370.0 °C | +/-0.007 % FS +/-0.01 °C +/-0.005 % FS +/-0.1 °C | |
| | Type J | FeCu-Ni | -90.00 °C to 900.00 °C | +/-0.005 % FS +/-0.01 °C | |
| | Type N | NiCrSi - NiSiMg | -90.00 °C to 999.99 °C 1000.0 °C to 1370.0 °C | +/-0.007 % FS +/-0.01 °C +/-0.005 % FS +/-0.1 °C | |
| | Type E | NiCr-CuNi | -90.00 °C to 700.00 °C | +/-0.005 % FS +/-0.01 °C | |
| | Type R | Pt13Rh - Pt | 0.00 °C to 999.99 °C 1000.0 °C to 1760.0 °C | +/-0.05 % FS +/-0.01 °C +/-0.03 % FS +/-0.1 °C | |
| | Type T | Cu-CuNi | -90.00 °C to 400.00 °C | +/-0.01 % FS +/-0.01 °C | |
| | Type B | Pt30Rh-Pt6Rh | 0.00 °C to 999.99 °C 1000.0 °C to 1820.0 °C | +/-0.05 % FS +/-0.01 °C +/-0.03 % FS +/-0.1 °C | |
| | Type S | Pt10Rh-Pt | 0.00 °C to 999.99 °C 1000.0 °C to 1760.0 °C | +/-0.05 % FS +/-0.01 °C +/-0.03 % FS +/-0.1 °C | |
| | Type L | Fe-CuNi | -90.00 °C to 900.00 °C | +/-0.005 % FS +/-0.01 °C | |
| | Type U | Cu-CuNi | -90.00 °C to 600.00 °C | +/-0.01 % FS +/-0.01 °C | |
| | Automatic comparison point compensation between 0 °C and 60 °C Accuracy of the comparison point Pt 100 DIN class A Possibility of connection through 4 mm connections free of thermal voltage and DIN and Mini DIN thermocouple connection | |
| | Standard signal input | |
| | Current (switchable) | mA | 0(4)...20 mA | +/-0.015 % FS +/-0.01 mA | |
| | Transmitter supply 24 VDC, I max = 30 mA Possibility of connection through 4 mm connections free of thermal voltage | |
| | | |
| | Automatic detection of an edge change, determining the hysteresis, Independent detection normally closed / normally open Possibility of connection through 4 mm connections free of thermal voltage Potential-free input contacts (Umax = 5 V, Imax = 1 mA) | |
| | Calibration reference sensor connection | |
| | Pt 100 | 4-wire | -90.00 °C to 850.00 °C | +/-0.005 % FS +/- 0.01 °C | |
| | Polynomial correctable through internal parameters or through external EEPROM in sensor Possibility of connection through 7-pin built-in socket | |
| | | | | | |