Catalogue Precision Temperature Calibrator
www.sika.net
print switch display
Page / 14
SIKA - 3626, 2255, 14826
/ 14
See other catalogues for SIKA
Text version of the page
Technical data
Resistance industrial thermometer according to DIN EN 60751
Pt 100 Pt 500 Pt 1000
2, 3, 4 wire
-90.00 °C to 850.00 °C
+/-0.005 % FS +/- 0.01 °C
Connection possibility through 4 mm connections free of thermal voltage
Thermocouples according to DIN EN 60584 / DIN 43710
Type K
NiCr-NiAl
-90.00 °C to 999.99 °C 1000.0 °C to 1370.0 °C
+/-0.007 % FS +/-0.01 °C +/-0.005 % FS +/-0.1 °C
Type J
FeCu-Ni
-90.00 °C to 900.00 °C
+/-0.005 % FS +/-0.01 °C
Type N
NiCrSi - NiSiMg
-90.00 °C to 999.99 °C 1000.0 °C to 1370.0 °C
+/-0.007 % FS +/-0.01 °C +/-0.005 % FS +/-0.1 °C
Type E
NiCr-CuNi
-90.00 °C to 700.00 °C
+/-0.005 % FS +/-0.01 °C
Type R
Pt13Rh - Pt
0.00 °C to 999.99 °C 1000.0 °C to 1760.0 °C
+/-0.05 % FS +/-0.01 °C +/-0.03 % FS +/-0.1 °C
Type T
Cu-CuNi
-90.00 °C to 400.00 °C
+/-0.01 % FS +/-0.01 °C
Type B
Pt30Rh-Pt6Rh
0.00 °C to 999.99 °C 1000.0 °C to 1820.0 °C
+/-0.05 % FS +/-0.01 °C +/-0.03 % FS +/-0.1 °C
Type S
Pt10Rh-Pt
0.00 °C to 999.99 °C 1000.0 °C to 1760.0 °C
+/-0.05 % FS +/-0.01 °C +/-0.03 % FS +/-0.1 °C
Type L
Fe-CuNi
-90.00 °C to 900.00 °C
+/-0.005 % FS +/-0.01 °C
Type U
Cu-CuNi
-90.00 °C to 600.00 °C
+/-0.01 % FS +/-0.01 °C
Automatic comparison point compensation between 0 °C and 60 °C Accuracy of the comparison point Pt 100 DIN class A
Possibility of connection through 4 mm connections free of thermal voltage and DIN and Mini DIN thermocouple connection
Standard signal input
Current (switchable)
mA
0(4)...20 mA
+/-0.015 % FS +/-0.01 mA
Transmitter supply 24 VDC, I max = 30 mA
Possibility of connection through 4 mm connections free of thermal voltage
Automatic detection of an edge change, determining the hysteresis, Independent detection normally closed / normally open Possibility of connection through 4 mm connections free of thermal voltage Potential-free input contacts (Umax = 5 V, Imax = 1 mA)
Calibration reference sensor connection
Pt 100
4-wire
-90.00 °C to 850.00 °C
+/-0.005 % FS +/- 0.01 °C
Polynomial correctable through internal parameters or through external EEPROM in sensor Possibility of connection through 7-pin built-in socket
-11-