Precision Temperature Calibrator - SIKA - #6

/ 14


catalogue search
P. 01
P. 02
P. 03
P. 04
P. 05
P. 06
P. 07
P. 08
P. 09
P. 10
P. 11
P. 12
P. 13
P. 14
Pages:


See other catalogues for SIKA

Text version of the page
The menu structure, convenient and simple to operate
select
7 KPM
Select mode
The Select mode offers the central setting and selection possibility for all processes specific to the calibrator.
Here you select your test pieces and decide for logger function or step program. Further selection possibilities such as display values and system settings are also possible.
calibr RTD calibr TC calibr mA switch test logger
step function display value system value info
adjustment
heating
13:24:56
switch test
-2 KPM
Switch test
You can perform here the automatic test of temperature switches and thermostats with self-detection of "normally open" or "normally closed".
The switching point with rising temperature Temp 1, the switching point with falling temperature Temp 2 and the automatically calculated hysteresis of the calibrators TP 38 000 are displayed.
block temp
79.50°C
set temp (Tmin)
50.00°C
ext. ref. 350.80 °C Temp 1 102.49 °C
Temp 2 97.11 °C
hysteresis 5.3 8 K
cooling
13:24:56
se^C^^^gger^^^TPMn
The calibrators TP 38 000 can be programmed for up to 8 completely automated calibrations. You make the settings of temperature steps, holding time and gradient directly at the calibrator. A PC download is possible.
The block and test piece temperatures are stored and are available for download. Reading off on the display is possible at any time.
You can perform automatic calibrations directly on site, without a PC or laptop, and the certificate can be produced later.
Select mode for test piece selection
With the aid of clearly laid out screen pages, you select in the Select mode all important parameters, e.g. temperature, supply, connection for the calibration of a wanted test piece.
Resistance thermometers RTD, thermocouples TC, transmitters and switches can be selected as test pieces.
The transmitter values are programmable. Switch on a transmitter supply is possible.
• ON OFF Seti CL
bloct temp set temp
101.10 199.05 302.70 402.00 499.80
100.00 200.00
400.00 500.00
date 2004/05/17 start 18:55:12
sensoi Pt100-4
heating
19:24:56
select mA
12 KPM
mA calibration • ON OFF
• T( 0mA) T(4mA)
0.00°C
T(20mA)
transmitter supply ON •OFF
connection
• 4mm-terminals
heating
15:17:40
select system 12 KPM
Select System mode
Here you can make general settings for the calibration system, such as selection of the languages, updating time and date etc.
• |Deutsch English
13:24:56 2003/03/17
brighness 100 %
contrast 100 %
alarm 610.00°C
switch off 620.00°C
working hors 7 h
work fac tor 14
heating
18:35:26
-6-
SIKA Dr.Siebert & Kuhn GmbH & Co. KG Struthweg 7-9 • 34260 Kaufungen • Germany • S.: +49 5605 803-0 • Fax: +49 5605 803-54 • e-mail: info@sika.net • www.SIKA.net

pageCatalog pdf di En 2012-02-07-15