SIKA Precision Measuring Instrument
... extends your calibration possibilities
Already integrated or easy to retrofit
Pi : -:■
The 4 mm connections for plug, cable shoes and open cable ends as well as a DIN and Mini DIN thermocouple connec­tion are available for connecting the test piece free of thermal voltages.
Resistance thermometers, thermocouples, temperature transmitters and switches must be operated in the cali­bration with a measuring instrument which measures and displays as temperature the output signals such as resistance values, thermal voltages and norm signals.
Use of our TP 38 000 with an integrated precision mea­suring instrument is unproblematic.
The measuring instrument can be retrofitted at any time at SIKA in an existing TP 38 000 E.
The switchable measuring input facilitates calibration, ad­justment and testing of:
• Resistance thermometers (RTD)
Pt 100, Pt 500 and Pt 1000 in 2, 3 or 4 wire
^ Thermocouples (TC) of the types K, J, N, E, R, T, B, S, L and U
^ Current signals 0(4)-20 mA of temperature transmit­ters (mA) with and without transmitter supply
^ Temperature switches and thermostats with "nor­mally open" and "normally closed" contacts
Calibration with external calibration reference sensors of­fers advantages especially for short probes, since both sensors are introduced at the same depth in the bore. The temperature of the external calibration reference sensor, of the test piece and the difference between the two appear on the display. Thus this calibration can also be performed simply.
The control functions are transfered on a external cali­bration reference sensor.
The measuring instrument is available in two versions:
• Retrofittable without difficulty in the TP 38 000 E ^ Available as desktop model TT Scan
The precision measuring instruments permit the simulta­neous connection of a test piece and of an external cali­bration reference sensor.