SITRANS F flowmeters
SITRANS F M
Transmitter MAGFLO MAG 5000/600
Application
For use with MAG 5000/6000 19" and MAG 1100 Ex ATEX/MAG 3100 Ex ATEX
Ex approval
MAG 1100 Ex [EEx e ia] IIB ATEX MAG 3100 Ex [EEx e ia] IIC ATEX
Cable parameter
Group Capacity in u.F Inductance in mH
IIC < 4.1 < 80
IIB < 45 < 87
IIA < 45 < 87
Ambient temperature
• During operation
• During storage
-20 ... +50 °C (-4 ... +122 °F) -20 ... +70 °C (-4 ... +158 °F)
• Material
• Width
• Height
• Rating
• Mechanical load
Standard 19" insert in aluminium/steel (DIN 41494) 21 TE (4.75")
3 HE (5.25")
IP20 / NEMA 2 to EN 60529 and DIN 40050
1 g, 1 ... 800 Hz sinusoidal in all directions to EN 60068-2-36
EMC performance
• Emission
• Immunity
EN 50081-1 (Light industry) EN 50082-2 (Industry)
Application
For use with professional transmitters MAG 5000 and 6000 19" to clean the electrodes on sensors MAG 1100 or MAG 3100
NB: Must not be used with intrinsically safe ATEX sensors
NB: Not to be used with sensors with Hastelloy and Tantalum electrodes
Cleaning voltage
AC cleaning DC cleaning
60 V AC 30 V DC
Cleaning period
60 s + 60 s pause period
Relay • Load
42 V/2 A
Operation
• Automatic
• Manual
Switch relay activated when cleaning is in progress Yes
No
Indicator lamps
LEDs: "ON" and "CLEANING"
Supply voltage and power consumption
115 ... 230 V AC, +10% ... -15%, 50 ... 60 Hz, 7 VA cleaning, 5 VA stand by 11 ... 30 V DC / 11 ... 24 V AC, 50 ... 60 Hz, 7 VA cleaning, 5 VA stand by
Ambient temperature
• During operation
• During storage
-20 ... +50 °C (-4 ... +122 °F) -20 ... +70 °C (-4 ... +158 °F)
Enclosure
• Material
• Width
• Height
• Rating
• Mechanical load
Standard 19" insert in aluminium/steel (DIN 41494)
21 TE (4.75") 3 HE (5.25")
IP20 / NEMA 2 to EN 60529 and DIN 40050
1 g, 1 ... 800 Hz sinusoidal in all directions to EN 60068-2-36
Electrode cleaning unit
Cleaning unit
The Siemens cleaning unit can be used with MAG 5000 or 6000 in 19" insert version.
The cleaning unit can be used in applications where the liner and subsequently the electrodes may be coated with deposits. If the coating is electrically insulating, the electrode signal will be reduced. If the coating is electrically inductive, the electrode sig­nal will be partly short-circuited and in both professional cases the accuracy of the meter will decrease (dependent on coating type and thick­ness).
Note:
The cleaning unit cannot be used for inflammable or explosive media!
Siemens FI 01 • 2008
DirectIndustry's Virtual Technical Library: PDF Catalogue | Technical Documentation | Brochure | Manual | Industrial directory | Specifications | Characteristics
Search Go
page 1 p.1
page 2 p.2
page 3 p.3
page 4 p.4
page 5 p.5
page 6 p.6
page 7 p.7
page 8 p.8
page 9 p.9
page 10 p.10
page 11 p.11
page 12 p.12
page 13 p.13
page 14 p.14
page 15 p.15
page 16 p.16
page 17 p.17
page 18 p.18
page 19 p.19
page 20 p.20
page 21 p.21
page 22 p.22
page 23 p.23
page 24 p.24
page 25 p.25
page 26 p.26
page 27 p.27
page 28 p.28
page 29 p.29
page 30 p.30
page 31 p.31
page 32 p.32
page 33 p.33
page 34 p.34
page 35 p.35
page 36 p.36
page 37 p.37
page 38 p.38
page 39 p.39
page 40 p.40
page 41 p.41
page 42 p.42
page 43 p.43
page 44 p.44
page 45 p.45
page 46 p.46
page 47 p.47
page 48 p.48
page 49 p.49
page 50 p.50
Pages:
1-50
51-100
101-150
151-200
201-250
251-300
301-350
351-388
pdf-page pdf di En 2008-10-41-10