LG200-DRD temperature transmitter DIN43650
12Pages

{{requestButtons}}

Catalog excerpts

LG200-DRD temperature transmitter DIN43650 - 1

SENSING TOMORROW LG200-DRD(H) Integrated thermal resistance temperature transmitter Product introduction LG200 integrated temperature transmitter adopts ASIC&SMT signal transmitting module, optional built-in backlight and button operation LCD display module. The integrated transient voltage terminal satisfy 4 grade standard (difference-mode voltage 2000V, common-mode voltage 4000V), suitable for bad surge voltage occasions. LG200 integrated temperature transmitter provides a flexible and reliable solution for any temperature measurement applications. LG200-DRD Flange installation LG200-DRH Thread installation Main parameters Temperature measurement Measuring medium The fluid which compatible with wetted parts Disclaimer: all the data used in the product description is not legally binding. Relevant technical details may be changed due to further improve

Open the catalog to page 1
LG200-DRD temperature transmitter DIN43650 - 2

SENSING TOMORROW LG200-DRD(H) Integrated thermal resistance temperature transmitter Technical Specifications Loading effects Measuring range and limit -50-400°C, min measuring range 100°C Above measurement range can be replaced by °F or K units. Provide other measuring range according to requirements. Adjust requirements: lower range value (LRV) and upper range value (URV) can be adjusted within the scope of the upper and lower range limit, smallest calibratable spans | URV-LRV | s upper range limit s±0.02% URL/100fi(refer to full scale output 20mA) Vibration effects Standard specifications...

Open the catalog to page 2
LG200-DRD temperature transmitter DIN43650 - 3

SENSING TOMORROW LG200-DRD(H) Integrated thermal resistance temperature transmitter Technical Specifications Damping time Environment condition Total damping time constant: equal to the sum of damping time of amplifer and sensor capsule Reaction time: £10s@ water flow 0.4m/s, outer diameter: 6mm Reaction time(Test standard: IEC60751, 10s@ water flow 0.4m/s) Tube diameter 10mm Tube wall thickness 1 mm Disclaimer: all the data used in the product description is not legally binding. Relevant technical details may be changed due to further improve

Open the catalog to page 3
LG200-DRD temperature transmitter DIN43650 - 4

SENSING TOMORROW LG200-DRD(H) Integrated thermal resistance temperature transmitter Technical Specifications Maximum medium flow rate(The maximum medium flow rate dured by thermal protection tube reduces with increases of insertion length, see chart below) Tube diameter 10mm Tube wall thickness 1mm Disclaimer: all the data used in the product description is not legally binding. Relevant technical details may be changed due to further improve

Open the catalog to page 4
LG200-DRD temperature transmitter DIN43650 - 5

SENSING TOMORROW LG200-DRD(H) Integrated thermal resistance temperature transmitter Product selection instruction Pressure sensor types Extension tube selection Electrical connection(voltage and current signal output) Extension tube length Process temperature The relation chart of thermal resistance terminal self-heating and process temperature Terminal temperature= environment temperature+ terminal self-healting Disclaimer: all the data used in the product description is not legally binding. Relevant technical details may be changed due to further improve

Open the catalog to page 5
LG200-DRD temperature transmitter DIN43650 - 6

SENSING TOMORROW LG200-DRD(H) Integrated thermal resistance temperature transmitterProduct selection instructionElectrical connection accessory (voltage and current signal output)Aviation plug straighter(JI) Aviation plug elbow (J2) label M12*1, 4-pins, aviation plug(H1), aviation plug straighter(J1, J4), aviation plug elbow (J2, J5) Brown White Blue Black Disclaimer: all the data used in the product description is not legally binding. Relevant technical details may be changed due to further improve

Open the catalog to page 6
LG200-DRD temperature transmitter DIN43650 - 7

SENSING TOMORROW LG200-DRD(H) Integrated thermal resistance temperature transmitter Product selection instruction Process connection select instruction Disclaimer: all the data used in the product description is not legally binding. Relevant technical details may be changed due to further improve

Open the catalog to page 7
LG200-DRD temperature transmitter DIN43650 - 8

LG200-DRD(H) Integrated thermal resistance temperature transmitter SENSING TOMORROW Product drawing and dimension Drawing and dimension (thread) with DIN43650 (D1) and without extension tube(unit: mm) Drawing and dimension (thread) with aviation plug (H1) and without extension tube( unit: mm) Drawing and dimension (thread) with DIN43650 (D1) and extension tube(unit: mm) Drawing and dimension (thread) with aviation plug (H1) and extension tube(unit: mm) Disclaimer: all the data used in the product description is not legally binding. Relevant technical details may be changed due to further...

Open the catalog to page 8
LG200-DRD temperature transmitter DIN43650 - 9

LG200-DRD(H) Integrated thermal resistance temperature transmitter SENSING TOMORROW Product drawing and dimension Drawing and dimension (tri-Clamp) with DIN43650 (D1) and extension tube ( unit:mm) Drawing and dimension (flange) with DIN43650 (D1) and extension tube (unit:mm) Disclaimer: all the data used in the product description is not legally binding. Relevant technical details may be changed due to further improve 2016.03.V1.0

Open the catalog to page 9
LG200-DRD temperature transmitter DIN43650 - 10

SENSING TOMORROW LG200-DRD(H) Integrated thermal resistance temperature transmitter Product drawing and dimension Disclaimer: all the data used in the product description is not legally binding. Relevant technical details may be changed due to further improve Aviation female plug straighter(J1)(unit: mm) Aviation female plug elbow(J2)(unit:mm) Aviation female plug straighter(J4) (unit: mm) Aviation female plug elbow(J5) (unit: mm)

Open the catalog to page 10
LG200-DRD temperature transmitter DIN43650 - 11

SENSING TOMORROW LG200-DRD(H) Integrated thermal resistance temperature transmitter Ordering information chapter Disclaimer: all the data used in the product description is not legally binding. Relevant technical details may be changed due to further improve

Open the catalog to page 11
LG200-DRD temperature transmitter DIN43650 - 12

SENSING TOMORROW LG200-DRD(H) Integrated thermal resistance temperature transmitter Ordering information chapter SENSING TOMORROW LEEC Shanghai LEEG Instruments Co.,Ltd ADD: No.100 Duhui Road, Minhang District, Shanghai China Postcode:201109 Tel: (86) 21-31261976 Fax: (86)21-31261975 E-mail:sales@leegsensor.com info@leegsensor.com Web: www.leegsensor.com Disclaimer: all the data used in the product description is not legally binding. Relevant technical details may be changed due to further improve

Open the catalog to page 12