DF-745 SGMax NanoTrace Manual - 94 Pages

Add to MyDirectIndustry favorites
DF-745 SGMax NanoTrace Manual

Catalog excerpts

Copyright 2011 by Servomex Corporation No part of this publication may be reproduced, stored in a retrieval system or transmitted in any form, or by any means including electronic, mechanical, photocopying, recording or otherwise without prior written permission of Servomex Corporation. NanoTrace Moisture Analyzer is a trademark of Servomex Corporation. VCR is a registered trademark of the Cajon Company

 Open the catalog to page 2

DF-745SG Operator Manual Firmware v5.3.8 Manual Version 110911 Your NanoTrace Moisture Analyzer has been designed, manufactured and is supported under ISO-9001 controls, thus helping to insure the highest possible standards of quality. Every analyzer that Servomex manufactures is tested and operated on a variety of gas concentrations to insure that it functions properly when you receive it. The certificate of calibration assures your analyzer has been calibrated on gases that are traceable to NIST standards. With proper maintenance, your analyzer should remain calibrated for years. For a...

 Open the catalog to page 3

Read Me First... Unpacking Procedure Follow the procedure below to unpack your NanoTrace Moisture Analyzer 1. Examine the condition of the packaging and its contents. If any damage is apparent, immediately notify the carrier and Servomex. Do not proceed with the installation. 2. Check the contents against the packing slip to make sure the shipment is complete. Unattached equipment may be shipped with the analyzer in supplemental packaging. Shortages should be reported to Servomex immediately. 3. All NanoTrace Moisture Analyzers are shipped with the following: 4. Open the analyzer door,...

 Open the catalog to page 5

Thank You Thank you for selecting the NanoTrace Moisture Analyzer. Servomex designs, manufactures, exhaustively tests, and supports every analyzer under the tightest quality controls. You should expect every Servomex analyzer to arrive in perfect working order and, with good maintenance, provide years of trouble-free service. Please call your nearest Servomex Business Center if you need assistance.

 Open the catalog to page 6

2 Cautions There are a number of warnings and cautions that must be observed to avoid damage to the analyzer as well to insure the safety of its users. The analyzer must be operated in a manner specified in this manual. Servomex cannot be responsible for direct or consequential damages that result from installing or operating the analyzer in a manner not described in this manual. Importantly, the analyzer has been designed for use with inert, non-toxic, non-combustible sample gases only. Servomex cannot be responsible for direct or consequential damages that result from using the analyzer...

 Open the catalog to page 13

CAUTION Do not setup or operate this analyzer without a complete understanding of the instructions in this manual. Do not connect this Analyzer to a power source until all signal and plumbing connections are made. CAUTION This analyzer must be operated in a manner consistent with its intended use and as specified in this manual. EMI DISCLAIMER This Analyzer generates and uses small amounts of radio frequency energy. There is no guarantee that interference to radio or television signals will not occur in a particular installation. If interference is experienced, turn-off the analyzer. If the...

 Open the catalog to page 14

3.1 Moisture Lowest Detection Level (LDL): 100 ppb @ Constant Conditions Resolution: Analytical (Smallest Detectable Change): 10 ppb Display: 0.1 ppb Accuracy: Greater of: 0-20 ppm range: ±5% of reading or ±25 ppb @ Constant Conditions 21-100 ppm range: ±10% of reading or ±100 ppb @ Constant Conditions Speed of Response: Typically less than 3 minutes to reach 90 percent of an upward step change Upset Recovery Time: Typically less than 5 minutes from a high ppb upset to within 10 ppb of the previously stable reading. Range: 0-100 ppm Background Gas Compatibility: All inert and passive gases...

 Open the catalog to page 15

Sample Temperature: 10° C to 80°C (50° F to 176° F) EMI Sensitivity: Tested to standards EN61000-3-3 and EN61326-1 Audible/Visual Alarm Status Indicators: Four Moisture levels, Temperature, Moisture cell Diagnostic, Zero Verification or Calibration in Process, Moisture Analyzer off-line, Analog output freeze during calibration. Alarm Relays: Four non-latching, independently assignable to moisture alarms or to moisture calibration-in-process indicator. SPDT contacts rated at 1 Amp @ 30 VDC. Fail safe action upon loss of power to alarm condition. Not designed to switch AC power. Storage...

 Open the catalog to page 16

Figure 1: Overall View

 Open the catalog to page 17

4 Installation, Start Up and Shut Down Installation of the analyzer requires the following steps be followed:  Connecting the pneumatic pressure service to the pneumatic inlet fitting Connecting the sample gas line to the analyzer inlet fitting Making the power connection to the analyzer POWER SWITCH Figure 2: Major Internal Components 4.1 Analyzer Installation 4.1.1 Pneumatic Pressure Line Connection - Optional The pneumatic gas connection (required for the optional gas panel) is a 1/8 inch compression fitting as shown in Figure 4 and requires 70 – 125 psig air or N2 pressure....

 Open the catalog to page 19

4.1.2 Sample Gas Connections 4.1.2.1 Sample Gas Inlet Connection Sample gas is connected to the analyzer via a ¼ inch male swivel VCR fitting labeled Process Inlet at the rear of the instrument as shown in Figure 3. Sample pressure of 30 – 150 psig is required and is regulated internally. Pre-purge the line by connecting to the analyzer (with a new VCR filter gasket) only finger tight and flowing gas for 15 minutes. Then tighten the inlet fitting. NOTE: A VCR filter gasket (supplied) should always be used to protect the gas delivery system from any particulate matter that may have collected...

 Open the catalog to page 20

4.1.2.2 Sample Gas Outlet Connection The sample gas outlet connection is a ¼ inch compression fitting labeled Moisture Sensor Outlet as shown in Figure 3. If a connection must be made to the exhaust, it must provide minimal backpressure. NOTE: If the Hydrogen Service Safety System is included, the sample outlet line must be made of steel. See page 83 for additional information. REGULATOR ADJUST BYPASS FLOW ADJUST PROCESS INLET PROCESS OUTLET BYPASS OUTLET Figure 4: Rear Gas Connections and Controls with Optional Gas Panel 4.1.3 Electrical Connections Open the door and locate the power...

 Open the catalog to page 21

All SERVOMEX catalogs and technical brochures

  1. 4900 Multigas

    4 Pages