4900 Multigas - 4 Pages

Add to MyDirectIndustry favorites
4900 Multigas

Catalog excerpts

HIGH PERFORMANCE CONTINOUS EMISSIONS (CEMS) ANALYZER, DESIGNED FOR MULTI-GAS MEASUREMENT OF CRITERION POLLUTANTS, GREENHOUSE GASES AND REFERENCE OXYGEN. FLEXIBLE SERVOPRO 4900 Multigas °{° The 4900 Multigas offers high specification multi-gas monitoring designed specifically for the continuous emissions monitoring of flue gas. This sensitive analyzer combines leading edge patented sensing technologies including paramagnetic O2, non-dispersive infrared (NDIR) techniques of Single Beam Single Wavelength (SBSW), and Gas Filter Correlation (GFx). When used with a suitable sampling system the 4900 Multigas provides unparalleled MCERTS/QUAL1 measurements of criterion pollutants NO, SO2, and CO, plus O2, and greenhouse gases CO2, N2O, and CH4. The 4900 Multigas combines this monitoring power with low-cost of ownership, delivering a highly attractive analytical package. With a compact and robust design, the small footprint of the 4900 Multigas requires minimum cabinet space and can be easily integrated into other systems, helping to reduce installation costs. Low operational requirements, facilitated through long calibration intervals by ultra-stable, high performance measurement technologies and auto-calibration capabilities, make the 4900 Multigas an ideal monitoring solution with an affordability to match its impressive monitoring performance. Provides a complete continuous emissions monitoring solution for flue gas analysis Ideal for criterion pollutant and greenhouse gas monitoring: % level O2, CO2 and CO, plus ppm level SO2, NO, CO, CH4 and N2O 2 4 Continuous multi-gas monitoring Digital communications for remote access: RS232/RS485 Modbus, PROFIBUS and Ethernet (Modbus TCP/ IP - to follow) EASY TO USE Small and compact: designed for simplified integration into existing systems and easy fit into a cabinet Auto-calibration facility Intuitive-use icon-driven color touchscreen for easy device interaction and configuration USB serial port for data logging and software upgrades LOW COST OF OWNERSHIP Reduced ongoing operational needs facilitated by auto-calibration function Extended calibration periods from ultra-stable, industry-leading Paramagnetic and SBSW IR and GFx IR sensing technologies UNRIVaLLEd PERFORmaNCE Non-depleting sensors for ■ Manufactured by Servomex - over ultra-stable, accurate and 60 years' experience innovating selective measurements and pioneering gas analysis and thousands of units used in the field every year Using the same reliable measurement technology as the SERVOPRO 4900, the new 4900 Multigas is backward-compatible with existing installations, providing trusted gas analysis while offering the advanced digital communications that many modern systems require. BENCHmaRk compliance In compliance with Low ■ Certified to MCERTS (EN 15627-3) Learn more about the SERVOPRO 4900 Multigas

 Open the catalog to page 1

PRODUCT OVERVIEW: 4900 Multigas A complete monitoring solution for CEMS gas analysis ALTERNATIVE PRODUCTS For industries and processes including power generation, petrochemical, refining, waste incineration, iron and steel, pulp and paper, and cement manufacture, continuous emissions monitoring is a regulatory requirement. The solution must be capable of offering the highest sensitivity and accuracy when dealing with multiple measurements for pollutants and greenhouse gases. No matter what your application needs, you’ll want a solution that’s easy to install and operate , while delivering...

 Open the catalog to page 2

TUNABLE DIODE LASER FLAME IONIZATION DETECTOR Whichever is the greater. ''Stated at a confidence interval of 95%

 Open the catalog to page 3

DEVICE SPECIFICATION 481.6mm (19") wide 544.2mm (21.4") deep ■ With expansion chassis, height is 265.5mm (10.5") Weight: ■ Main unit: approx 14kg (30.9lb) ■ Expansion chassis: approx 13.7kg (30.2lb) (dependent on number and type of sensors used) Operating Temperature: ■ 5oC - 45oC (41 oF - 113oF) Warm Up Time: ■ Warm up time is typically 24 hours from cold start at 20oC (68oF), may be longer for the higher sensitivity measurements. Power Requirements: Sample Gas Connections: ■ Sample inlet is 1/8" NPT female ■ Sample outlet is 1/4" NPT female Certifications: ■ EN 61010-1: Safety...

 Open the catalog to page 4