4.3 Connecting Hose
Sensirion recommends a hose with an inner diameter of 1/8 to 3/20 inch (3.18 to 3.8mm). Due to the dynamic measurement principle, a small air flow is required (Figure 6) which
leads to a dependence on the length of the hose (Figure 7). Tubes up to 1 m show less than 1 % error of the measured value (Figure 7).
« 350 -
CO
° 300 -
Length of the connecting hose
Deviation of Measured Value
0.5 m (20 inch)
- 0.4 %
1.0 m (40 inch)
- 0.8 %
2.0 m (80 inch)
-1.6%
4.0 m (160 inch)
- 3.2 %
5 150 4
a. 100 -------------------------------\ \
I 50 -------------i
<=> 0 Y^--\-\-
0 30 60 90
Mass flow through SDP1108 [seem] Figure 6: Typical air flow through the SDP1108. Figure 7: Influence ofthe length ofthe connecting hose on
Please note: 1 scc/min = 1 cm3/min at 0°C and the accuracy (using 3/16 inch inner diameter).
1013 mbar pressure (1 sccm = 0.001 norm liter). Example: a 50 Pa difference pressure is shown as
49.8 Pa when using 0.5 m tube with 3/16 inch diameter.
www.sensirion.com
SDP1108 - preliminary data sheet v0.95 February 2007
9/12