Application Note Relative Humidity Calculation
Calculation of relative humidity at a given temperature
RH
1
, T
1
, D
1
RH
2
, T
2
, D
2
A
Figure 1: Chamber >
A
with constant dew-point >
D1=D2
. It is assumed that the dew-point in chamber >
A
is constant, i.e. >
D
1 =D
2 . Given the relative humidity >
RH
1 at temperature >
T
1 in chamber >
A
, the relative humidity >
RH
2 at temperature >
T
2 can be calculated using >
n 12
(1)
with parameters >
eeRHRH
TTnmTTTmT
2211 .For calculations with micro-controllers, formula (1) can be expressed as >
RH2 = RH1 * 10^( 7.6523 * T1 / ( 243.12 + T1 )) / 10^( 7.6523 * T2 / ( 243.12 + T2 )) For further information please refer to the Application note “Dew point calculation”. >
www.sensirion.comSensirion