Sartorius WDS 400Selective detection of surface water, capillary water and water of crystallization
Thermogravimetry Water, not moisture >
StDry nitrogenOven up to 400°CNitrogen + H 2 OP 2 O 5 + H 2 O cellDry nitrogen with H Thermogravimetric methods, such as the
oven drying method, use the weight loss
of a sample to determine the total con-tent of all volatile components and not,however, the pure water content. As a
rule, the latter task is performed using
electrochemical techniques that arebased on the principle of coulometry(coulomb = electric charge). The most
commonly known methods are coulo-
metric Karl Fisher titration for solid andliquid samples and the phosphorus pen-
toxide method for trace analysis of gases.
However, both methods require compli-
cated equipment; moreover, KF titration
necessitates the use of additional chemi-
cals in order to perform an analysis. TheWDS 400 Water Detection System from
Sartorius combines these three standardmethods into a high-resolution and easy
procedure for selective detection ofwater in solids and pastes. >
2 + O 2 r r r Electronics ▲▲▲▼ Backweighing r
lr r Balance PCPrinter Get all three in one Highly accurate and selective The WDS 400 adopts the principle of
convection heating from the oven drying
method in order to drive out the entiremoisture from a sample. A ceramic disccoated with extremely hygroscopic phos-
phorus pentoxide P2O5 completely
absorbs the water from the resulting gasmixture and bonds water molecules tophosphoric acid H3PO4 on the disc sur-
face in a chemical reaction. By coulome-
try, i.e., by an electric current generatedat the ceramic disc, phosphoric acid isbroken down into phosphorus pentoxide
P2O5, hydrogen H and oxygen O. Based
on Faraday’s law, it is known how muchcurrent is necessary in order to split off
all hydrogen atoms from a chemicalcompound. Thus, the WDS 400 uses the
amount of electric current to calculatethe quantity of water driven out of asample. 18 This combination method works so accu-
rately that it is even possible to detect
one single microgram of water. Beyondthat, the WDS 400 enables water frac-tions to be differentiated according to
surface water, capillary water and water
of crystallization (the latter is chemicallybound water). Easy operation All the user has to do is just weigh-in asample. The WDS 400 does not require
any complicated handling of detectionreagents, many of which are toxic. For measurement of the water content,the user can choose the type of carriergas, either nitrogen (Class 5.0) or room
air. For using room air, the WDS 400 has
a built-in pump and a
drying unit. >
Coulometry Microwave resonance NIR spectroscopy