1. Catalogs
  2. Emerson Automation Solutions - ROSEMOUNT
  3. Rosemount 3051S Series of Instrumentation

Rosemount 3051S Series of Instrumentation

Rosemount  3051S Series of Instrumentation
1 / 162 PagesView full catalog

Rosemount 3051S Series of Instrumentation

Product catalog summary
Overview
The Rosemount 3051S Series offers scalable solutions for pressure, flow, and level measurements, enhancing production, quality, energy efficiency, and safety. It reduces process variability, improves process insight, minimizes maintenance, and complies with regulatory demands.
Rosemount 3051S SuperModule™ Platform
This platform provides advanced measurement capabilities with high reliability and accuracy, featuring a hermetic SST design, ultra performance accuracy, and a 15-year stability and warranty. It is SIL3 capable and certified for safety instrumented systems.
Product Variants
  • Coplanar Transmitter: Integrated manifold and seal system solutions with dual-capacitance sensor technology.
  • In-line Transmitter: Direct threaded connections and piezoresistive sensor technology.
  • MultiVariable Transmitter: Combines multiple measurements for accurate flow readings and customizable compensation.
  • DP Flowmeters: Integrates with primary elements for a complete flowmeter assembly, reducing installation costs.
  • ERS System: Provides stable DP Level measurements with digital architecture.
  • Level Transmitter: Combines pressure transmitters with direct-mount seals for easy installation.
Advanced Functionality
The series supports WirelessHART capabilities, reducing installation time and costs. It offers advanced diagnostics for process and safety systems, and remote display and interface options for enhanced accessibility.
Specifications and Certifications
Detailed specifications and certifications ensure compliance with industry standards. The document includes ordering information, dimensional drawings, and accessory options.
Specifications and Options
  • Diagnostic Functionality: Options include FOUNDATION Fieldbus and Advanced HART diagnostics.
  • Mounting Brackets: Various options for mounting brackets are available.
  • Software Configuration: Custom software configurations are available.
  • Alarm Limits: NAMUR alarm and saturation levels are available in high and low configurations.
  • Hardware Adjustments: Options for zero, span, alarm, and security adjustments.
  • Custody Transfer: Measurement Canada Accuracy Approval is available.
  • Product Certifications: A wide range of certifications including ATEX, FM, CSA, IECEx, INMETRO, and others.
  • Shipboard Approvals: Approvals from American Bureau of Shipping, Bureau Veritas, Det Norske Veritas, and Lloyds Register.
  • Sensor Fill Fluid: Inert sensor fill fluid is available.
  • Display Type: Options for PlantWeb LCD display and remote mount LCD displays.
  • Pressure Testing and Cleaning: Hydrostatic testing and special cleaning services are available.
  • Calibration and Material Traceability: Calibration certificates and material traceability certifications per EN 10204 3.1 are offered.
  • Quality and Safety Certifications: Prior-use certificates and safety certifications to IEC 61508 are available.
  • Transient Protection and Drinking Water Approval: Transient terminal block and NSF drinking water approval options are provided.
MultiVariable Transmitter
  • Performance and Measurement: Offers superior flow calculations with options for gas, natural gas, steam, and liquid measurements.
  • Measurement Types: Includes differential pressure, static pressure, and temperature measurements.
  • Pressure and Temperature Ranges: Various ranges for differential and static pressure, as well as temperature inputs, are specified.
  • Isolating Diaphragm and Process Connections: Multiple materials and configurations for isolating diaphragms and process connections are available.
Conclusion
The document provides comprehensive ordering information for the Rosemount 3051S series, highlighting the flexibility and customization options available to meet specific application needs.
Specifications and Compliance
The document outlines the compliance of wetted materials with NACE MR0175/ISO 15156 for sour oil field environments and NACE MR0103 for sour refining environments. A NACE certificate can be obtained by ordering with Q15 or Q25. The pressure range should be specified based on maximum static pressure, not differential pressure. Tantalum diaphragm material is available only with pressure sensor/module code G. Silicone fill fluid is standard, and environmental limits apply to certain materials.
Model and Configuration Details
The Rosemount 3051SAM and 3051SAL transmitters are designed for ERS applications with integrated transmitters and direct mount seals. The 3051SAL model offers scalable advanced level transmitters with options for ultra and classic performance classes. Various pressure module types and ranges are available, including coplanar and in-line gage and absolute types. The document provides detailed ordering information, including model numbers, housing styles, seal system types, and fill fluids.
Options and Accessories
Extended product warranties and electronic remote sensor connection cables are available. Software configuration options include custom configurations and gage pressure calibration. Alarm limits and ground screw assemblies are specified for enhanced safety and functionality. A wide range of product certifications is available, including ATEX, FM, CSA, IECEx, and INMETRO.
Special Procedures and Certifications
Special procedures include hydrostatic testing and cleaning for specific services. Certifications available include calibration certificates, material traceability, and compliance to NACE standards.
Key Considerations
The document emphasizes the importance of specifying product materials and options by the purchaser. Detailed specifications and additional information can be found in the referenced pages and manuals.
Overview
The document provides detailed information on the Rosemount 3051S Series Scalable Level Transmitters, highlighting their features, specifications, and ordering information. These transmitters combine high-performance pressure measurement with durable diaphragm seals.
Key Features
  • Variety of process connections including flanged, threaded, and hygienic seals.
  • Quantified performance for the entire transmitter/seal assembly (QZ option).
  • Supports HART, FOUNDATION Fieldbus, and wireless protocols.
Specifications Detailed specifications are available on page 103, including performance classes, configuration types, pressure ranges, and transmitter outputs.
Ordering Information The document provides a comprehensive guide to ordering the Rosemount 3051SAL Scalable Level Transmitter, including:
  • Performance classes with accuracy and warranty details.
  • Configuration types for liquid level measurement.
  • Pressure module types and ranges.
  • Transmitter output options including HART, Fieldbus, and wireless.
  • Housing styles and materials.
  • Seal system types and connection options.
  • Capillary lengths and seal fill fluids with temperature limits.
Product Certifications Certifications and standards are detailed on page 127, ensuring compliance with industry requirements.
Additional Options The document outlines additional options such as wireless configurations, extended warranties, diagnostic functionalities, and mounting brackets.
Recommendations The document suggests selecting starred options for best delivery and notes that expanded offerings may require additional lead time.
Overview
The document provides detailed technical specifications and operational guidelines for the Rosemount 3051S Series, a line of pressure transmitters used in industrial applications. Key features include zero and span adjustment, power supply requirements, load limitations, and advanced diagnostics.
Specifications
  • Zero and Span Adjustment: Can be set anywhere within the range, with span needing to be greater than or equal to the minimum span.
  • Output: Two-wire 4–20 mA output, user-selectable for linear or square root output, with digital process variable superimposed on the signal.
  • Power Supply: External power supply required, with specific voltage requirements depending on the model and options.
  • Load Limitations: Maximum loop resistance is determined by the external power supply voltage.
Advanced Diagnostics
  • Statistical Process Monitoring (SPM): Provides statistical data to detect process anomalies.
  • Power Advisory Diagnostic: Detects and notifies of degraded electrical loop integrity.
FOUNDATION Fieldbus
  • Power Supply: Requires external power supply; operates on 9.0 to 32.0 Vdc.
  • Function Blocks: Includes resource, transducer, LCD display, analog input, PID with auto-tune, and more.
  • Diagnostics Suite: Includes SPM and Plugged Impulse Line Detection.
WirelessHART (IEC 62591)
  • Output: Operates at 2.4 GHz DSSS with various antenna options.
  • Local Display: Optional seven-digit LCD display for user-selectable information.
Environmental and Operational Limits
  • Temperature Limits: Ambient and storage temperature ranges specified, with variations for LCD display and wireless output.
  • Pressure Limits: Detailed maximum working and burst pressure limits for different sensor modules.
  • Humidity and Damping: Operates at 0–100% relative humidity with user-selectable damping time.
Additional Features
  • Software Upgrades: Easy field upgrades using FOUNDATION Fieldbus Common Device Software Download procedure.
  • PlantWeb Alerts: Provides diagnostic alerts and recommendations for instrumentation issues.
Temperature Specifications The document outlines various temperature limits for different materials and configurations. For example, Silicone 704 has an upper temperature limit of 401 °F (205 °C) without a direct-mount extension, which increases with extensions. Other materials like Silicone 705 and UltraTherm 805 have higher temperature thresholds.
Transmitter Turn-on Time The document specifies the turn-on times for different transmitter models, such as 2 seconds for the 3051S model and 5 seconds for diagnostics.
Failure Mode Alarm The document describes the alarm configurations for transmitter failures, including high and low alarm signals that are software or hardware selectable. Safety-certified transmitter failure values are also provided.
Material Selection Emerson offers a variety of materials for Rosemount products, emphasizing the purchaser's responsibility to analyze process parameters when selecting materials.
Electrical and Process Connections Various connection types are detailed, including NPT, G1/2, and M20 conduit connections, as well as process connections for different sensor modules.
Product Certifications The document lists certifications for the Rosemount 3051S series, including European Directive Information, and certifications for use in North America and Canada, such as FM Explosionproof and CSA Intrinsically Safe certifications.
Overview
The document provides detailed technical specifications, certifications, and safety conditions for the Rosemount 3051S Series transmitters. It includes information on various international standards and certifications, such as ATEX, IECEx, INMETRO, and Chinese standards, highlighting the specific conditions for safe use in hazardous environments.
Specifications The transmitters are designed to operate under specific voltage, current, and power conditions. The document lists different models with their respective electrical parameters, including voltage (Ui), current (Ii), power (Pi), capacitance (Ci), and inductance (Li).
Certifications and Standards The document references multiple standards, including EN 60079-0, EN 60079-11, IEC 60079-0, IEC 60079-11, and others. Each certification, such as ATEX, IECEx, and INMETRO, has specific markings and conditions for safe use, indicating the equipment's suitability for different hazardous zones and temperature classes.
Special Conditions for Safe Use Key conditions include ensuring proper ingress protection (IP20 to IP66), using appropriate cable entries and blanking plugs, and considering the inability of certain models to withstand a 500 V insulation test. The document emphasizes the importance of following manufacturer instructions for installation and maintenance to ensure safety.
Temperature Classifications The transmitters are classified into temperature classes T6, T5, and T4, with corresponding process temperature ranges. These classifications are crucial for determining the suitability of the equipment in various environmental conditions.
Installation and Maintenance Recommendations The document advises on the use of protective measures, such as polyurethane paint finishes for aluminum enclosures, and highlights the need for secure installation to maintain ingress protection. It also notes the importance of using certified components and adhering to specific standards during installation and maintenance.
International Considerations The document includes specific guidelines for different regions, such as Brazil and China, detailing the applicable standards and special conditions for safe use in those areas.
Specifications and Compliance
The document outlines the compliance of the product with various international standards for explosive gas atmospheres, including IEC60079-27:2008 for FISCO field devices. It specifies the use of shielded cables grounded in non-hazardous areas and prohibits end users from altering internal components.
Installation and Maintenance Standards
Installation and maintenance must adhere to standards such as GB3836.13-1997, GB3836.15-2000, and GB3836.16-2006, which cover repair, electrical installations, and inspection in explosive gas atmospheres.
Certification and Markings
The product holds various certifications across different regions, including EAC, FM, CSA, and ATEX, with specific markings indicating compliance with intrinsic safety and explosion-proof standards. Special conditions for safe use are noted, such as temperature ranges and material considerations.
Energy Limiting Parameters
Details are provided on maximum input voltage, current, power, and internal parameters for different models, ensuring compatibility with intrinsically safe circuits.
International Certifications
The document lists certifications from countries like the USA, Canada, Japan, and China, each with specific standards and conditions for safe use. It also includes combinations of certifications for broader compliance.
Telecommunication Compliance
Wireless devices require certification to adhere to RF spectrum regulations, with Emerson ensuring compliance to avoid legal issues.
Installation in North America
Guidelines for using Division and Zone marked equipment in accordance with the US NEC and Canadian CEC are provided, ensuring suitability for area classification and temperature class.
Overview
The document provides detailed technical specifications, certifications, and safety conditions for the Rosemount 3051S Series, a pressure measurement device used in various industrial applications. It includes information on standards, markings, and special conditions for safe use across different regions and certifications.
Specifications and Standards The device complies with multiple international standards, including CAN/CSA, EN, IEC, and ABNT NBR IEC standards. It is designed for use in hazardous environments, with certifications for flameproof, intrinsic safety, and dust ignition-proof applications.
Markings and Certifications The device is marked for various hazardous location classifications, such as Class I, Division 1, and Zone 0, with temperature classes ranging from T6 to T4. It holds certifications from ATEX, IECEx, INMETRO, and other regional bodies, ensuring compliance with safety and performance standards.
Special Conditions for Safe Use The document outlines specific conditions for safe installation and maintenance, such as the need for protective measures against environmental impacts, the use of certified cable entries and blanking plugs, and adherence to manufacturer instructions to ensure device longevity and safety.
Electrical Parameters Detailed electrical parameters are provided for different configurations, including HART, FOUNDATION Fieldbus, and FISCO. These include maximum input and output voltages, currents, power, capacitance, and inductance values.
Regional Certifications The device is certified for use in various regions, including Europe, Brazil, China, and Korea, with specific conditions and standards applicable to each region. It also includes combinations of certifications for broader applicability.
Additional Certifications The Rosemount 3051S Series holds additional certifications from maritime and offshore classification societies such as ABS, BV, DNV, and Lloyds Register, indicating its suitability for marine and offshore applications.
Conclusion The Rosemount 3051S Series is a versatile and highly certified pressure measurement device suitable for use in hazardous environments across multiple industries and regions. Its compliance with international standards and certifications ensures reliability and safety in various applications.
Overview
The document provides detailed dimensional drawings and specifications for the Rosemount 3051S Series, a range of transmitters and flowmeters used in industrial automation. It includes various configurations and mounting options, along with accessory information.
Dimensional Drawings The document includes figures illustrating the dimensions of different transmitter configurations, such as those with coplanar sensor modules, traditional flanges, in-line sensor modules, and various mounting configurations. Dimensions are provided in both inches and millimeters.
Transmitter Configurations Several configurations are detailed, including:
  • Coplanar Sensor Module with Flange
  • In-line Sensor Module
  • High Gain, Remote Mount Antenna
  • Various mounting configurations (pipe, panel, remote display)
Flowmeter Specifications The document details the dimensions and specifications for different flowmeter models, including the Annubar Flowmeter, Compact Orifice Flowmeter, and Integral Orifice Flowmeter. Tables provide specific dimensional data for each model.
Level Transmitter Configurations Various scalable level transmitter configurations are described, including those with FF, EF, RF, RT, SS, and SC seals. Each configuration includes detailed dimensions.
Accessories The document outlines the Rosemount Engineering Assistant software packages, which support flow configuration for the 3051S series. It includes system requirements and options for modems and connecting cables.
Contact Information The document provides contact details for Emerson Process Management's global headquarters and regional offices across North America, Latin America, Europe, Asia Pacific, and the Middle East and Africa.
Trademarks and Legal The document concludes with a list of trademarks and legal disclaimers related to the products and software mentioned.
See more

Catalog excerpts

Rosemount  3051S Series of Instrumentation-1

Product Data Sheet January 2016 00813-0100-4801, Rev UA Rosemount™ 3051S Series of Instrumentation Scalable pressure, flow, and level solutions Innovation reaching across your operation With the Rosemount 3051S Series of Instrumentation, operations can be optimized in these critical areas: production, quality, energy efficiency, and safety and environment. By leveraging the power of the scalable Rosemount 3051S across the entire operation, you’ll be able to minimize process variability, gain greater process insight, reduce maintenance and downtime, and meet regulatory demands. What’s more, it’s easy to use, ensuring the full potential of the measurement investment is real

 Open the catalog to page 1
Rosemount  3051S Series of Instrumentation-2

Rosemount 3051S SuperModule™ Platform The most advanced pressure, flow, and level measurements  The all-welded hermetic SST design delivers the industry's highest field reliability Ultra performance provides up to ±0.025% accuracy and 200:1 rangedown Ultra for Flow performance provides up to ±0.04% of reading and 14:1 flow turndown 15-year stability and 15-year limited warranty SIL3 Capable: IEC 61508 certified by an accredited 3rd party agency for use in safety instrumented systems up to SIL 3 (minimum requirement of single use [1oo1] for SIL 2 and redundant use [1oo2] for SIL 3) Rosemount...

 Open the catalog to page 2
Rosemount  3051S Series of Instrumentation-3

Rosemount 3051S MultiVariable™ Transmitter See ordering information on page 21.  Combines differential pressure, static pressure, and process temperature measurements along with mass and energy flow in a single device Compensates for 25+ different variables providing accurate and repeatable flow readings Customize pressure and temperature compensation for any flow application Easily configure flow and device parameters with Engineering Assistant Software Rosemount 3051SF DP Flowmeters See ordering information on page 29.  Integrates the 3051S with Rosemount’s industry leading primary elements...

 Open the catalog to page 3
Rosemount  3051S Series of Instrumentation-4

Advanced functionality WirelessHART® (IEC 62591) capabilities Available on coplanar, in-line, multivariable, DP flowmeters and level transmitters  Quickly deploy new pressure, level and flow measurements in 70% less time Eliminate wiring design and construction complexities to lower costs by 40 - 60% Reduce pipe penetrations and impulse piping with industry-leading multivariable technology Extended range antenna capabilities provide access to remote locations Delivering over a decade of maintenance free performance with 15-year stability and 10-year power module life Advanced diagnostic capabilities...

 Open the catalog to page 4
Rosemount  3051S Series of Instrumentation-5

Rosemount 3051S Coplanar Pressure Transmitter Rosemount 3051S Coplanar Pressure Transmitters are the industry leader for differential, gage, and absolute pressure measurement. The coplanar platform allows seamless integration with manifolds, primary elements, and seal solutions. Capabilities include:  Ultra, Ultra for Flow, and Classic Performance 4-20 mA HART, Wireless, FOUNDATION™ Fieldbus protocols Safety Certification (Option Code QT) Advanced Diagnostics (Option Code DA2) Remote Display and Interface (Option Code M7, M8, or M9) 3051S Coplanar Pressure Transmitter Additional Information...

 Open the catalog to page 5
Rosemount  3051S Series of Instrumentation-6

Table 1. Rosemount 3051S Scalable™ Coplanar Pressure Transmitter Ordering Information ★ The Standard offering represents the most common options. The starred options (★) should be selected for best delivery. __The Expanded offering is subject to additional delivery lead time. Gold-plated Alloy 400 (includes graphite-filled PTFE O-ring) Materials of construction Size Process connection Flange material Drain vent None (no process flange) Assemble to Rosemount 305 integral manifold Assemble to Rosemount 304 or AMF manifold and SST traditional flange Assemble to Rosemount 304 or AMF manifold to SST...

 Open the catalog to page 6
Rosemount  3051S Series of Instrumentation-7

Table 1. Rosemount 3051S Scalable™ Coplanar Pressure Transmitter Ordering Information ★ The Standard offering represents the most common options. The starred options (★) should be selected for best delivery. __The Expanded offering is subject to additional delivery lead time. Coplanar flange Coplanar flange Coplanar flange Coplanar flange Coplanar flange Coplanar flange Traditional flange Traditional flange Traditional flange Traditional flange Traditional flange Traditional flange Traditional flange Traditional flange DIN-compliant traditional flange Vertical mount level flange Vertical mount...

 Open the catalog to page 7
Rosemount  3051S Series of Instrumentation-8

Table 1. Rosemount 3051S Scalable™ Coplanar Pressure Transmitter Ordering Information ★ The Standard offering represents the most common options. The starred options (★) should be selected for best delivery. __The Expanded offering is subject to additional delivery lead time. PlantWeb housing PlantWeb housing Wireless PlantWeb housing Wireless PlantWeb housing Junction Box housing with output for remote display and interface Junction Box housing with output for remote display and interface Junction Box housing with output for remote display and interface Quick Connect (A size Mini, 4-pin male...

 Open the catalog to page 8
Rosemount  3051S Series of Instrumentation-9

Table 1. Rosemount 3051S Scalable™ Coplanar Pressure Transmitter Ordering Information ★ The Standard offering represents the most common options. The starred options (★) should be selected for best delivery. __The Expanded offering is subject to additional delivery lead time. FOUNDATION Fieldbus diagnostics suite Advanced HART diagnostics suite PlantWeb enhanced measurement functionality(15)(17) H01 FOUNDATION Fieldbus fully compensated mass flow block Coplanar flange bracket, all SST, 2-in. pipe and panel Traditional flange bracket, CS, 2-in. pipe Traditional flange bracket, CS, panel Traditional...

 Open the catalog to page 9
Rosemount  3051S Series of Instrumentation-10

Table 1. Rosemount 3051S Scalable™ Coplanar Pressure Transmitter Ordering Information ★ The Standard offering represents the most common options. The starred options (★) should be selected for best delivery. __The Expanded offering is subject to additional delivery lead time. Delete transmitter drain/vent valves (install plugs) SST coplanar flange without drain/vent ports ATEX Flameproof ATEX Intrinsic Safety ATEX FISCO Intrinsic Safety (FOUNDATION fieldbus protocol only) ATEX Flameproof, Intrinsic Safety, Type n, Dust ATEX Dust TIIS Flameproof TIIS Intrinsic Safety FM Explosion-proof, Dust Ignition-proof...

 Open the catalog to page 10

All Emerson Automation Solutions - ROSEMOUNT catalogs and technical brochures

  1. FRL Accessories

    14  Pages

  2. TankRadar Pro

    32  Pages

  3. Roxar PTPT15K

    2  Pages

  4. SW-300

    4  Pages

  5. DL8000

    12  Pages

  6. AMS Suite

    16  Pages

  7. Line Card

    2  Pages

*Prices are pre-tax. They exclude delivery charges and customs duties and do not include additional charges for installation or activation options. Prices are indicative only and may vary by country, with changes to the cost of raw materials and exchange rates.