Laser Optical Micrometer, RF656 Series

Request ("Master") - 01 h, 83h

Message ("Master") - 89h, 80h, 80h, 83h

and for lower byte, code of parameter - 08h, value of parameter - 39h

Request ("Master") - 01 h, 83h

Message ("Master") - 88h, 80h, 89h, 83h

14. Parameterization program

14.1. Function

The RF65X-SP software (fwww.riflek.com/resource/files/rf65x-sp-.zip) is in-

tended for:

1) Testing and demonstration of work of RF65x series sensors;

2} Setting of the sensor parameters;

3) Reception and gathering of the sensor data signals

14.2. Obtaining connection to micrometer

Once the program is started, the pop-up window emerges:

COM Port/ USB :

UART Bond rale

To obtain connection it is necessary:

• select COM-port whereto the sensor is connected (logical port if the sensor is

connected via USB-adapter)

• select transmission rate (Baud rate) at which the sensor will work

• select the sensor network address, if necessary

• press the Connect button.

If the selected parameters correspond to the parameters of the micrometer inter-

face, the program will identify the micrometer, read and display its configuration parame-

ters:

RF656 [Revision 3.0110th October 2011


pageCatalog pdf di En 2012-05-22-31