RaySafe DXR+

RaySafe DXR+

RaySafe DXR+

Product catalog summary
Introduction
The RaySafe DXR+ is a compact tool designed for aligning X-ray and light fields, ensuring compliance with international regulations that recommend alignment within 2% of the Source-to-Image Distance (SID) or ±5 mm at the chest wall for mammography. It provides objective, reproducible, and immediate read-outs down to 30 kVp, making it suitable for both radiographic and mammographic equipment.
Specifications
  • Range: ±5 cm (±2 in)
  • Segment Sensors: 41 solid state sensors with corresponding LCD display segments
  • Segment Resolution: 0.25 cm (0.1 in)
  • Power: Automatic on when exposed, automatic off after 1 minute of inactivity
  • Battery Life: 6 – 8 years (CR1632), based on 2000 exposures per year
  • Operating Temperature: 10 – 40°C (50 – 104°F)
  • Storage Temperature: -20 – +60°C (-4 – +140°F)
  • Size: 15 x 30 x 145 mm (0.59 x 1.18 x 5.71 in)
  • Weight: 75 gr (2.6 oz)
Procedures
  1. Positioning: Place the RaySafe DXR+ within the light field, aligning the center line with the light field edge.
  2. Exposure: Expose the device to X-ray or light.
  3. Read Out: Check the display for any deviation in the X-ray or light field.
Features
  • Automatic power on and off
  • No need for light field adjustment to a square phantom before exposure
  • Eliminates the need for film development
Recommended Generator Settings
  • Mammography: kVp Max 50, mA >100, SID <65 cm
  • Radiographic: kVp Max 70-100, mA >50-200, SID <100 cm
  • Exposure Time: >10 ms
  • No added tube filtration
Patents
  • Swedish Patent: 526928
  • American Patent: 7313223
See more

Catalog excerpts

RaySafe DXR+-1

RaySafe DXR+ An easy tool for X-ray/Light field alignment

 Open the catalog to page 1
RaySafe DXR+-2

RaySafe DXR+ The RaySafe DXR+, a Direct X-ray Ruler, represents the latest in today’s technology for alignment of the light and radiation field. (Traditionally screenfilm cassettes have been used for X-ray and light field alignment measurements. The digitalization of X-ray imaging makes this method obsolete as cassettes and film processors are often unavailable.) It is the ideal solution for performing most accurate measurements for radiography and mammography. Actual size Quick checks The RaySafe DXR+ is an easy-to-use instrument and the ideal solution for improving digital imaging in the radiographic...

 Open the catalog to page 2
RaySafe DXR+-3

How it works 1. Position the RaySafe DXR+ Position the -symbol within the light field. Align the center line with the light field edge. The light field edge is aligned with the yellow center line. Light field 2. Expose Expose the RaySafe DXR+ to X-ray or light. X-ray field The X-ray field is 1.2 cm (0.5 in) inside of the light field. Adjust the collimator. Light field 3. read out Read any X-ray or light field deviation in the RaySafe DXR+ display.

 Open the catalog to page 3
RaySafe DXR+-4

Recommended generator settings Segment sensors/ read out 41 solid state sensors and corresponding LCD display segments 0.25 cm (0.1 in) Segment resolution All segments should turn on when the RaySafe DXR+ is fully irradiated. Automatic when exposed Operating temperature Storage temperature Battery life time Automatic after 1 min of inactivity Function test Exposure time: >10 ms No added tube filtration Unfors Instruments has changed its name to Unfors RaySafe www.raysafe.com All specifications may change without notice. Swedish patent 526928. American patent 7313223.

 Open the catalog to page 4

All RaySafe catalogs and technical brochures

  1. RaySafe i3

    8  Pages

  2. RaySafe x2

    8  Pages

  3. RaySafe ThinX

    3  Pages

  4. RaySafe Solo

    5  Pages

  5. RaySafe Xi

    8  Pages

Archived catalogs

  1. Unfors Xi

    4  Pages

  2. Mammographic

    4  Pages

*Prices are pre-tax. They exclude delivery charges and customs duties and do not include additional charges for installation or activation options. Prices are indicative only and may vary by country, with changes to the cost of raw materials and exchange rates.