PM-1200 Series Imaging Photometer and Colorimeter - Radiant Imaging - #2

/ 2


catalogue search
P. 01
P. 02
Pages:


See other catalogues for Radiant Imaging

Text version of the page

PM-1200 Color Series

CCD Imaging Photometer, Radiometer, and Colorimeter

TYPICAL SPECIFICATIONS* DEFECT DETECTION

2-DimensionalMeasurement CapabilitiesUnits CCD Resolution (px) LuminanceRadiance Illuminance Irradiance Luminous Intensity Radiant Intensity CIE Chromaticity Coordinates Correlated Color Temperature (CCT)

(advanced capabilities) (a (a dv dv an an ce ce d d ca ca pa pa bi bi li li ti ti es es ) )

Footlambert, Cd/cm
2 , Cd/m
2 , Nit, Mnit, mnitW/sr/m
2 , W/sr/ft
2 , W/sr/cm
2 , mW/sr/m
2 Footcandles, Lux, mLux, Mlux, Lux-Sec W/m
2 , W/ft
2 , W/m
2 , mW/m
2 , MW/m
2 , W-Sec/m
2 Candela W/sr CIE (x,y) and (u’,v’) Kelvin (CCT)1392x1040 ± 3% CCD Camera A/DDynamic Range 12 bits = 4,096 grayscale levels Luminance Range
0.05 nit minimum 10
10
nit maximum with ND filters
SystemAccuracy IlluminanceLuminance (Y)Chromaticity Coordinates (x,y) ± 3%
1 1 ± 0.005
1 Short TermRepeatability IlluminanceLuminance Chromaticity Coordinates (x,y) ± 0.5%
2 ± 0.5% Interface USB 2.0
2 ± 0.001
2 Minimum ColorMeasurement Timefor 100 cd/m 5 seconds
2 Camera Field ofView (FOV) 2 to 32 degrees Dimensions: 242mm (H) x 154mm (W) x 76mm (D) Weight: 5.5 lbs. (2.5 kg.) * Specifications subject to change without notice. Operatingtemperature: 0 - 30 C
o Operating humidity: 20-70 % non-condensing 1. Based on Illuminant A, D 65, or user calibration for specific spectra. Based on a virtual detector size of 100 pixels. Specification is for every point within the field of view of the camera.2. At every point within the field of view of the camera, based on virtual detector size of 100 pixels.

SYSTEM REQUIREMENTS

Pentium IV, 1.3 GHz Processor or faster, 1 GB RAM Windows® 2000 or XPUSB 2.0 Interface 15321 Main St. NE, Duvall, Washington, 98019, USPhone: (425) 844-0152 Fax: (425) 844-0153