Digital Mass Flow Controller / Meter
AFC 202D / AFM 302D
AFC & AFM Series
Specifications
Flow Range (equivalent N2) 30 slm to 250 slm
up to 400 slm with H2
Control Range 2 and 100% F.S.
Valve Type Electromagnetic
Valve Rest Position Normally Open or Closed
Linearity ±0.2% of setpoint (per calibration gas)
Repeatability ±0.15% of F.S.
Sensibility to Mounting Position ±0.1% of F.S.
Step Response Time ..2.5 sec. typical (SEMI E17-91)
Temperature Range 5 — 50°C, non-condensing
Temperature Coefficient <0. 5% F.S. /°C
Maximum Inlet Pressure 1 MPa (10 bar, 150 psi)
Minimum Differential Pressure 150 kPa for 30 slm F.S.
200 kPa for 50 slm F.S.
250 kPa for 100 slm F.S.
300 kPa for 200 slm F.S.
450 kPa for 400 slm F.S.
Pressure Coefficient <0.1% F.S./105 Pa
Wetted Materials 316 L Stainless steel
Kel-F
Seals material
Surface Finish 0.4 ěm (16 ěinch) Ra max
Leak Integrity <2.10-9 atm.scc/sec (He)
Standard Seals Viton
Neoprene
Fittings 3/8’’ M VCR, Swagelok, other on request
Weight 3.7Kg
Input and Output
Mass Flow Controller ±15 VDC, 1 A
Mass Flow Meter ±15 VDC, 50 mA
Input and Output Signal from 0 to 5 VDC
Digital Control RS232C, Active full time
Electrical Connector Sub-D 15 pins Male
Features
For high flow from 30 slm up to 250 slm, the highest performances of digital P.C.B. combined with
elastomeric seals on the Qualiflow Therm AFC 202D and AFM 302D :
- Full potential of digital technology.
- 100% compatible with analog AFC202 using Sub-D 15 pin adapter (P/N Q2001733-09 or Q2001733-10).
- Special pressure compensated stainless steel valve for ultra fast flow control.
Accuracy
During calibration, a polynomial curve (fourth degree) calculated
from 6 points gives an accuracy of :
±1% of setpoint if setpoint >30% of F.S.
±0.3% of F.S. if setpoint <30% of F.S.
Less Inventory
Up to 10 calibration curves of different gases can be stored into the memory, easily changeable by user. To keep the benefit of accuracy, the maximum factor between the full scales is 3. A factor of 5 maintains an accuracy less than ±1%, compatible with analog MFCs.
Optimized Numerical Control
The control of gas flow is done by a numerical algorithm assuming a control without overshoot at any setpoint and improving the response time. Each calibration curve is stored with its own regulation parameters.
Communication Modes
Analog mode : the MFC is 100% compatible with analog series, with the advantage to communicate via serial RS232C for maintenance, calibration.
RS485 option : allows to control up to 32 MFCs under MODBUS protocol (RJ11 connector).
PROFIBUS high speed fieldbus.
Calibration
MFCs are calibrated with Nitrogen
Reference Standards Temperature .. 273.15K
Pressure .. 101325 Pa
Options
External sensor input
DeviceNet or Profibus fieldbus connection
RS485 / Modbus with 2 RJ11 connections
|
|