| UNIT DESIGN Multi patient unit - Designed to supply 4-6 dialysis machines, depending on dialysis set up and machine type. 5pm pre-filter - Protects reverse osmosis (RO) membrane. Integral tank - Protects against feedwater pressure fluctuations and allows easy cleaning. Integral boost pump - Enables RO membranes to operate at optimum performance. RO membrane - Removes up to 98% of dissolved minerals, and up to 99% of organic, bacterial and particulate contaminants. Permeate divert on start-up - Ensures consistent permeate quality. Audible alarm with mute facility - Provides an alarm monitoring permeate quality, cleaning chemical contamination and over pressure. User friendly display - Back-lit display clearly shows information concerning the unit operation in both text and graphic formats. Semi-automatic chemical clean - With automatic chemical draw, recirculation and rinse, cleaning is straight forward and trouble free. Standby recirculation for use in ITU / Acute applications - Ensures the bacterial quality is maintained so that the unit is ready for immediate use. Flexible start/stop facility - Interface to allow remote start/stop of the unit either via the dialysis machine, a programmable timer, or via a remote key switch. RS232 Interface - Provides easy connection to a remote PC. Point of use recirculation - The design of the system means that water continually recirculates to the point of use, ensuring the bacterial quality of the water is maintained. |