PT TF Transit Time Ultrasonic Flow Meter
Technology and Specifications
Principles of Operations T°P View of PiPe
Transit time flow meters utilize two transducers which function as both ultrasonic transmitters and receivers. The flow meters operate by alternately transmitting and receiving a frequency modulated burst of sound energy between the two transducers. The burst is first transmitted in the direction of fluid flow and then against fluid flow. Since sound energy in a moving liquid is carried faster when it travels in the direction of fluid flow (downstream) than it does when it travels against fluid flow (upstream), a differential in the times of flight will occur. The sound's time of flight is accurately measured in both directions and the difference in time of flight calculated. The liquid velocity (V) inside the pipe can be related to the difference in time of flight (dt) through the following equation:
V = K*D*dt, where K is a constant and D is the distance between the transducers.
Specifications
Reflect-Mode Configuration Transit Time
DESCRIPTION
SPECIFICATION
Liquid Types
Most clean liquids or liquids containing up to 40% suspended solids or aeration
Power
11-30 VDC @ 0.25 A
Velocity
0.1 to 40 FPS (0.03 to 12.4 MPS)
Inputs/Outputs
4-20 mA Output (standard)
Resolution 12-bit for all outputs
Power Source
Insertion loss 5 V maximum
Loop impedance 900 Ohms maximum
Isolation Can share ground common with power supply - isolated from piping system
Turbine Frequency Output / TTL-Pulse Output (standard)
Switch selectable
Type Non-ground referenced AC / Ground referenced square wave
Amplitude 100 mV peak to peak minimum / 5 VDC
Frequency range 0 - 1,000 Hz
Duty cycle 50% ±10%
Display
2 line x 8 character LCD; 0.7", 7 segment top row; 0.35", 14 segment bottom row
Rate: 8 maximum digits, lead zero blanking
Total: 8 maximum digits, exponential multipliers from -1 to +6
Units Engineering: feet, gallons, fts, million-gal, barrels (liquor & oil), acre-feet, lbs, meters, ms, liters, million-liters, kg
Rate: sec, min, hr, day
Ambient Temperature
General purpose: 0 °F to +185 °F (-20 °C to +85 °C)
Hazardous locations: 0 °F to 105 °F (-20 °C to +40 °C)
Liquid Temperature
0 °F to +185 °F (-20 °C to +85 °C)
Enclosure
NEMA 3 (Type 3) ABS or polycarbonate
Transducer
CPVC, Ultem®, brass or stainless steel hardware
Line Sizes
1/2" to 100" (12 mm to 2540 mm)
Line Materials
Carbon steel and stainless steel ANSI pipe; copper and plastic tubing
Accuracy
±1% of reading at rates above 1 FPS (0.3 MPS)
±0.01 FPS (0.003 MPS) of reading at rates lower than 1 FPS (0.3 MPS)
Response Time
0.3-30 seconds, adjustable
Protection
Reverse-polarity, surge suppression
Approvals
General requirements: ANSI/ISA 582-01
Hazardous locations: ANSI/ISA 12.12.01
Class I Div. 2, Groups C & D (Integral systems only)
CSA C22.2 No. 213, E79-15-95 (PTTF3 and PTTF4 models only)
UltraLink Utility
Windows® software utility, requires serial communication cable and Windows® 95, 98, 2000 or XP