Gas/Liquefied Gas Sampling Inlet System ensures reproducible injection of gases and liquefied petroleum gases (LPG). Controlled Rate Sample Drive provides reproducible, controlled rate of sample introduction into analyzers by syringe or sample boat.

ANTEK 9000 TOTAL SULFUR AND NITROGEN ANALYZERS

9000NS - TOTAL NITROGEN & SULFUR ANALYZER - 9054 Fast, precise; one injection – two results

Model 734
PRODUCT SPECIFICATIONS Detection Method Analytical Range* 9000N Nitrogen Analyzer: Pyro-chemiluminescence™ 9000S Sulfur Analyzer: Pyro-fluorescence™ 9000N Nitrogen Analyzer: ppb to 17%9000S Sulfur Analyzer: ppb to 40% Precision ±40 ppb or ±2% RSD, whichever is greater Typical Sample Size* Typical Analysis Time Gas 10 cc LPG 15 µl Liquid to 20 µl Solid: up to 500 mg*Analytical range and sample size may be a function of sample matrix and method of introduction
Model 735
Temperatures Gases / Liquids: less than one minute Solids: 2 - 15 minutes (dependant upon sample type/weight) Combustion zone: ambient to 1100°C in 1°C increments Power 115
Model 737
Temperature Programmable Inlet enables analysis of solid materials and viscous liquids up to 500 mg; ideal for additives in poyolefins (slip, antioxidants, etc.), thermoplastics, resins and coal. Multi-Matrix Sampler
VAC 50/60 Hz 12 Amp230 boat inlet system allows introduction of solids and liquids; thermoelectrically cooled chamber speeds time between sample runs. Robotic Liquid Autosampler Dimensions & Weight
VAC 50/60 Hz 6 Amp Gases robotically operated, syringe-based, vertical injection sampler for use with 9000 and Multi-Matrix or Boat Inlet; handles 105 (optional 200) samples. 51.5 cm h (20.276”) x 47.0 cm w (18.504”) x 60.0 cm d (23.622”) 39.5 kg (87 lb) Dry oxygen: 300-600 cc/min 99.75%, 3 bar (40 psig), >5 ppm moisture Dry carrier (argon or helium): 50-200 cc/min 99.99%, 3 bar (40 psig), > 5 ppm moisture
Model 740
ORDERING SPECIFICATIONS
for addition to Model 9000S Sulfur Analyzer. Sulfur Detector Kit Model 9000S Sulfur Analyzer designed for rapid analysis of solid, liquid and gas samples for total chemically bound sulfur. Accessories include a pyrolysis tube, 10ml syringe, operation/service manual, fittings and gas line tubing required for system operation. Desktop or laptop computer included. Complies with ASTM D5453 and D6667. Model 9000N Nitrogen Analyzer designed for rapid analysis of solid, liquid and gas samples for total chemically bound nitrogen. Accessories include a pyrolysis tube, 10ml syringe, operation/service manual, fittings and gas line tubing required for system operation. Desktop or laptop computer included. Complies with ASTM D4629, D5176, D5762 and D6069. Model 9000NS Nitrogen/Sulfur Analyzer designed for rapid analysis of solid, liquid and gas samples for simultaneous detection of total chemically bound nitrogen and sulfur. Accessories include a pyrolysis tube, 10ml syringe, operation/service manual, fittings and gas line tubing required for system operation. Desktop or laptop computer included. Complies with ASTM D4629, D5176, D5453, D5762, D6069, and D6667. Nitrogen Detector Kit for addition to Model 9000N Nitrogen Analyzer. Technology used in ANTEK’s laboratory instrumentation is also available for process/on-line applications.
Model 748
U.S. Patents: #4018562, #4066409, #4351801, #4352779, #4678756, #4778764, #4843016; U.K. Patent: #1578281; F.R.G. Patent: #P3116049.2 Due to continuing product development, specifications subject to change at any time without notice. Germany Badstrasse 3-5, P.O. Box 1241 D-97912 Lauda-Königshofen

FOR ADDITIONAL INFORMATION

YOUR LOCAL REPRESENTATIVE: USA France 8824 Fallbrook Drive, Houston, Texas 77064 Phone: 800.444.TEST [281.580.0339] | Fax: 281.580.0719 sales@paclp.com | service-lab@paclp.com BP 70285 - VERSON - 14653 CARPIQUET Cedex +33 (0) 231 264 300 | fax +33 (0) 231 266 293 sales@paclp.fr | service@paclp.fr +49 9343.6400 | fax +49 9343.640.101 sales@paclp.de | service@paclp.de Singapore 10, Eunos Road 8, #12-06 Singapore Post Centre 408600 +65 6742 8453 | fax +65 6742 8759 sales@paclp.com.sg | service@paclp.com.sg
© Copyright 2006 PAC L.P.

|

9000NS ver 1.0

|

Printed in USA 0606 www.paclp.com