ANTEK 9000 TOTAL SULFUR AND NITROGEN ANALYZERS
Fast, precise; one injection – two results
Due to continuing product development, specifications subject to change at any time without notice.
PRODUCT SPECIFICATIONS
Detection Method
9000N Nitrogen Analyzer: Pyro-chemiluminescence™
9000S Sulfur Analyzer: Pyro-fluorescence™
Analytical Range*
9000N Nitrogen Analyzer: ppb to 17%
9000S Sulfur Analyzer: ppb to 40%
Precision
±40 ppb or ±2% RSD, whichever is greater
Typical Sample Size*
Gas 10 cc
LPG 15 ìl
Liquid to 20 ìl
Solid: up to 500 mg
*Analytical range and sample size may be a function of sample matrix
and method of introduction
Typical Analysis Time
Gases / Liquids: less than one minute
Solids: 2 - 15 minutes (dependant upon sample type/weight)
Temperatures
Combustion zone: ambient to 1100°C in 1°C increments
Power
115 vac 50/60 Hz 12 Amp
230 vac 50/60 Hz 6 Amp
Gases
Dry oxygen: 300-600 cc/min 99.75%, 3 bar (40 psig),
>5 ppm moisture
Dry carrier (argon or helium): 50-200 cc/min 99.99%,
3 bar (40 psig), > 5 ppm moisture
Dimensions & Weight
51.5 cm h (20.276”) x 47.0 cm w (18.504”) x 60.0 cm d (23.622”)
39.5 kg (87 lb)
ORDERING SPECIFICATIONS
Model 9000S Sulfur Analyzer designed for rapid analysis of solid, liquid and gas
samples for total chemically bound sulfur. Accessories include a pyrolysis tube, 10ml
syringe, operation/service manual, fittings and gas line tubing required for system
operation. Desktop or laptop computer included. Complies with ASTM D5453 and
D6667.
Model 9000N Nitrogen Analyzer designed for rapid analysis of solid, liquid and gas
samples for total chemically bound nitrogen. Accessories include a pyrolysis tube,
10ml syringe, operation/service manual, fittings and gas line tubing required for
system operation. Desktop or laptop computer included. Complies with ASTM D4629,
D5176, D5762 and D6069.
Model 9000NS Nitrogen/Sulfur Analyzer designed for rapid analysis of solid, liquid
and gas samples for simultaneous detection of total chemically bound nitrogen and
sulfur. Accessories include a pyrolysis tube, 10ml syringe, operation/service manual,
fittings and gas line tubing required for system operation. Desktop or laptop computer
included. Complies with ASTM D4629, D5176, D5453, D5762, D6069, and D6667.
Nitrogen Detector Kit for addition to Model 9000S Sulfur Analyzer.
Sulfur Detector Kit for addition to Model 9000N Nitrogen Analyzer.
Technology used in ANTEK’s laboratory instrumentation is also available for process/
on-line applications.
U.S. Patents: #4018562, #4066409, #4351801, #4352779, #4678756,
#4778764, #4843016; U.K. Patent: #1578281; F.R.G. Patent: #P3116049.2
Model 734
Model 735
Model 737
Model 748
Model 740
© Copyright 2006 PAC L.P. | 9000NS ver 1.0 | Printed in USA 0606
Temperature Programmable Inlet
enables analysis of solid materials
and viscous liquids up to 500 mg;
ideal for additives in poyolefins
(slip, antioxidants, etc.),
thermoplastics, resins and coal.
Multi-Matrix Sampler
boat inlet system allows
introduction of solids and
liquids; thermoelectrically
cooled chamber speeds time
between sample runs.
Robotic Liquid Autosampler
robotically operated, syringe-based,
vertical injection sampler for use
with 9000 and Multi-Matrix or
Boat Inlet; handles 105 (optional
200) samples.
Controlled Rate Sample Drive
provides reproducible,
controlled rate of sample
introduction into analyzers by
syringe or sample boat.
Gas/Liquefied Gas Sampling
Inlet System
ensures reproducible injection
of gases and liquefied
petroleum gases (LPG).
|
|