ORTEC® GammaScreen-8™ 8-Channel Rapid Analysis Screening of Multiple Gamma Emitting Radionuclides in Clinical Samples Easy to Use Rapid and Simple Gamma Radionuclide Screening • A very compact solution for rapid screening of hundreds of samples/day for possible radioactive contamination. • No sample processing: Pass/fail evaluations against analysis limits. • 700 samples may be counted in 20 hrs of operation. • MDA (LOD)1 = 52 Bq/L, 300 second count. • 8 samples (10 ml or 50 ml) counted simultaneously with automated data entry by bar code scanning (optional). • 8 completely independent NaI(Tl) well detectors and count chains. • Digitally stable electronics. • Full spectral data retained for detailed analysis. 1Minimum Detectable Activity or Limit of Detection. ORTEC Gross Gamma Counter Sample Evaluation Counting System #1 Sample Date/Time: Friday, October 16 2009 11:13 Actual (CPS) Limit (CPS) Evaluation SampleID: 09-911-000064-1GG Detector: 1 6.36 4.05 FAILED SampleID: 09-911-000065-1GG Detector: 2 3.87 3.98 PASSED SampleID: 09-911-000066-1GG Detector: 3 3.43 3.80 PASSED SampleID: 09-911-000067-1GG Detector: 4 3.54 3.84 PASSED SampleID: 09-911-000068-1GG Detector: 5 6.44 3.82 FAILED SampleID: 09-911-000069-1GG Detector: 6 3.43 3.68 PASSED SampleID: 09-911-000070-1GG Detector: 7 3.55 11.59 Passed SampleID: 09-911-000071-1GG Detector: 8 3.58 11.00 Passed