Technical Data
Sensor AF16-EX-N
Material: Material:
measuring cell made of stainless steel SS 316 Ti, 1.4571 (standard)
Special materials:
SS 316 L (1.4435), 1.4539, 1.4462, TFM 4215, Hastelloy®C4, Hastelloy®C22, Titanium, Tantalum, Monel®400, Inconel® 625, PR and others on request.
Line size:
№ to 8", (DN 6 to DN 200) Process connections:
ASME Flange, DIN Flange, JIS Flange, Tri-Clamp, BBS-Clamp, Female Thread NPT, Female Thread DIN ISO 228/1 G, SanitaryThread (DIN 11851), and others on request.
Gaskets:
Viton®, EPDM (FDA), EPDM (USP Class VI), Kalrez®, Chemraz®, Fluoraz®, Buna (NBR), Silicone, Viton®/FEP (FDA), and others on request.
Windows:
Pyrex®, Sapphire
Optical path length:
1 mm - 1000 mm Process pressure:
10 mbarto 325 bar, (0.15 psi to 4713 psi), depending on process connection, material and design
Temperature specification:
elevated or reduced ambient temperatures may require restrictions to the operating temperatures.
ambient temperature: -30 °C to +40 °C, (-22 °F to +104 °F)
process temperature: -30 °C to +120 °C, (-22 °F to +248 °F)
periodic 15 min/day (no hazardous environment):
-30 °C to +150 °C, (-22 °F to +392 °F)
periodic 30 min/day (no hazardous environment):
-30 °C to +140 °C, (-22 °F to +284 °F)
duringtransport: -20 °C to +70 °C, (-4 °Fto +158 °F)
AirPurge:
connectors available as standard
Light source:
incandescent tungsten lamp: 5.0 V DC, 775 mA, typical lamp life 3to5 years
Wave length range:
730 nm - 970 nm
Detector:
silicon photo diode, hermetically sealed Calibration:
basic calibration in CU (concentration units)
Measuring range:
any measuring range between 0 - 0.05 to 5 CU
Resolution:
< ± 0.05 % of respective measuring range Repeatability:
< ± 0.5 % of respective measuring range Linearity:
specific to application, <±l%of respective measuring range Protection:
all optical parts protected accordingto NEMA 4X (IP65)
Cable Specification:
Fixed cable connection 2m, both side
0 - 400 m, (0 - 1312 ft.) appr. IIC T6/T5 (Groups A,B,C,D)
401 - 1000 m, (1313 - 3280 ft) appr. IIB T6/T5 (Groups C,D)
Hazardous locations:
II 2G EEx ia IIC/IIBT6/T5
Class I, Div. 1, Groups A, B, C, D
Approval report:
DMT ATEX E176 FMGJ.I. 3013884
Certificates:
ISO 9001:2000, ATEX, FM, PED, CE, HP0
Use with C4000 Converters
Options
AF16-EX-HT-N
high temperature ex-proof model perman.:
-30°Cto+240°C, (-22°Fto+464°F)
periodic 15 min/day:
-30 °C to +260 °C, (-22 °F to +500 °F)
periodic 30 min/day:
-30 °C to +250 °C, (-22 °F to +482 °F)
Validation adapter
modularadapterwith application specific validation filter for sensor verification
B~ fl W*i
" >■ i i
Measuring cells for any application
optek-Danulat GmbH ~~
Emscherbruchallee 2 45356 Essen Germany
Phone: +49-(0)201-63409-0 Fax: +49-(0)201-63409-999 E-Mail: info@optek.de Homepage: www.optek.com
optek-Danulat bv
Grote Brugse Grintweg 12 4005 AH Tiel Netherlands
Phone: +31-(0)344-683-800 Fax: +31-(0)344-653-950 E-Mail: info@optek.nl Homepage: www.optek.com
optek-Danulat, Inc.
N 118 W 18748 Bunsen Drive Germantown, Wl 53022, USA Phone: +1-262-437-3600 Fax: +1-262-437-3699 Toll free: +1-800-371-4288 E-Mail: info@optek.com Homepage: www.optek.com