Differential Scanning Calorimetry (DSC) - F1 SeriesThermogravimetry (TG) - F1 Series

F1 Iris TG 209 F1 Iris
® The TG 209 F1 Iris
® is a vacuum-tight,top-loading thermo-microbalance with direct temperature measurement at the sample crucible. The water-cooled micro- industrial furnace allows fast heating and cooling for a high sample throughput. The TG 209 Proteus A unique feature of the TG 209 F1 Iris
® is the ability to couple it to both a massspectrometer (MS) and a FTIR spectro me-ter, either individually or simultaneously – even when equipped with the ASC.•Temperature range: 10 ... 1000°C•Weighing range: 2000 mg •Resolution: 0.1µg •
® Auto-Vac can be equippedwith the Automatic Sample Changer (ASC) for 64 crucibles. Unstable samples can be loaded in sealed crucibles, which are opened by the piercing device just before starting a measurement. Sample specific measurements and evaluations are fully controlled by a flexible macro in the innovative 32-bit
® -ready•Vacuum: 10
-2 TG 209 F1 Iris ® mbar•3 mass flow controllers • Cooling speed <15 min (1000°C-100°C)
® ® ® •Unique c-DTA and SuperRes software. software•Same ASC as for DSC 204 F1 Phoenix
® DSC 204 F1 Phoenix
® The premium DSC 204 Ï„- Sensor(high resolution) F1 Phoenix
® in-tegrates a number of new hardware and software techniques which can be used in application fields such as polymers, foods, pharmaceuticals, inorganics, metals, composite materials, and so on.•Temperature range: -180 ... 700°C•Detection limit: 0.1µW •Choice of sensors: µ -Sensor (high sensitivity)•Unique gas-tight DSC furnace ensures purity of measuring atmosphere•3 integrated mass flow meters forgas control and recording•Cooling systems: compressed air,intracooler (down to -85°C) and LN
2 /GN
2 BeFlat (down to -180°C)•UV extension for photo calorimetry •Unique
® and ISSP
® techniques•Temperature modulation •hv/vis extension for photocalorimetry•Upgradable with Automatic SampleChanger (ASC) for up to 64 samples and references
DSC 204 F1 Phoenix ® with ASC

5