GHP 456 Titan™ - NETZSCH-Gerätebau - #9

/ 12


catalogue search
P. 01
P. 02
P. 03
P. 04
P. 05
P. 06
P. 07
P. 08
P. 09
P. 10
P. 11
P. 12
Pages:


See other catalogues for NETZSCH-Gerätebau

Text version of the page

Impact of temperature difference

Titan In case of any imbalance in thesystem (inhomogeneities in the plate temperatures, imperfect gap control or relevant radial heat losses), differ-ent temperature differences cause deviations in the results. The GHP 456
® does not show anyimpact of the temperature difference used for the tests and the resulting thermal conductivity. This is simple proof of the nearly perfect system design.

Application: PUR foam

Insulation of modern house roofs,cryo-tanks or even ships requires materials featuring both low thermal conductivity and high mechanicalstability. Polyurethane (PUR) foamsoffer these properties. Presented here is a comparison of a test with an HFM at room temperature and a GHP test down to -160°C. Both results agree perfectly. Additionally, the GHP result shows the impact ofcell-gas condensation between -50°C and -125°C.