DSC 404 F3 Pegasus
® >
– Applications
In this example, two differentsamples of the same aluminum alloy(Al-Mg-Si) were measured with the
DSC 404 F3 Pegasu s >
® during heatingand cooling. The melting and solid-ification of the alloy are clearlyvisible in the measurement results.The differences between the twosamples, however, are minimal. Thecharacteristic temperatures (onsets,peaks) all agree within 0.3 K. The
differences between the peak areasare less than 1% for both heatingsand coolings. The good agreementbetween the two measurementresults demonstrates the excellent
performance and reproducibility ofthe DSC 404 F3 Pegasus >
® .The specific heat flow rate of ironwas measured between room tem-perature and 1620°C using a DSC404 >
Phase transitions of iron
Reproducible testing ofaluminum alloy
F3 Pegasus >
® equipped with aRhodium
industrial furnace. The peak at770°C is due to a change in the
magnetic properties of the material(the Curie transition). At peaktemperatures of 926°C and 1399°C,two structural changes (changes inthe crystal structure) occurred. Most
likely due to impurities in the mate-rial, these temperatures are slightlyshifted from the literature values forpure iron. Melting occurred at1534°C (extrapolated onset). The
heat of fusion was 266.1 J/g. This isless than a 1.5% deviation fromtypical literature values for pure iron. >
8