Catalogue DSC 404 F1 Pegasus® - product brochure
www.netzsch-thermal-analysis.com
print switch display
Page / 12
Contact the
Manufacturer
Where to buy
this product ?
Request
a Quote
NETZSCH-Gerätebau -
/ 12
See other catalogues for NETZSCH-Gerätebau
Text version of the page

DSC 404 F1 Pegasus

®

– Applications

Specific heat accuracy ofalumina (Al

2

O

3

)

Presented in the figure is a specificheat test result for a polycrystallinealumina sample between room tem- perature and 1600°C. Additionallyshown are literature values for purealumina. It can clearly be seen thatthere are no significant differencesbetween the literature values andthe test results. The maximum devia-tions are in the range of 2%, whichdemonstrates the outstanding per-formance of the DSC 404 F1 Pegasus
® .

Reproducibility of a diopsideglass powder

Presented here are two measure-ment results for diopside glass pow-der. The glass transition occurredbetween 723°C and 745°C.Crystallization was measured at 883°C (onset temperature). Meltingtook place at 1390°C (main peaktemperature). The characteristictemperatures and the correspondingenthalpy changes are in good agree- ment between the two differentsamples. Furthermore, the specificheat itself doesn’t show relevantdifferences. This indicates the ex-cellent stability and reproducibility of the DSC 404 F1 Pegasus
® system.