You may also be interested in
Calorimeter, Tester, Test set, Mass spectrometer, Thermal conductivity analyzer
Text version of the page
DIL 402 C
The Vacuum-Tight Pushrod Dilatometer DIL 402 C –
The Perfect Instrument for all Demanding Measurement Tasks
The DIL 402 C is the result of 50 years´
worth of experience in building
dilatometers at NETZSCH.
tube
vacuum
flange
sample
carrier
sample
thermocouple
pushrod
The DIL 402 C has an Invar measurement system with a high-resolution
measurement transducer and comprehensive thermostatic control, offering
the highest degree of accuracy,
reproducibility and long-term stability
for application temperatures up to
2000°C.
The instrument´s horizontal design
with motorized pushrod and easy-tomove furnace makes it simple to insert
samples, even when sample geometries
are less than ideal. A thermocouple in
direct proximity to the sample enables
reproducible temperature measurement.
This allows for use of the c-DTA®
software to calculate endothermic and
exothermic effects in the sample.
Thus, in addition to determining
characteristic expansion values, it is
also possible to gather additional information about phase transitions, solidstate reactions or chemical reactions
such as oxidation.
4
sample
heating
element
This premium instrument has been
optimized for application in the
traditional areas of ceramics and glass
as well as for research in fields such
as advanced ceramics or powder
metallurgy.
shut-off valve
(gas outlet)
reversing
tube
DIL 402 C – 3-dimensional
sectional view
front
panel
furnance
flow-through
protective tube
At a glance
The DIL 402 C is characterized by its:
Wide temperature range
(-180°C to 2000°C via
exchangeable furnaces)
Large sample dimensions
(length: up to 50 mm,
diameter: up to 19 mm)
Adjustable contact pressure
(15 cN to 45 cN)
Extremely low drift, high
resolution
Vacuum-tight design
Capability of coupling to gas
analysis systems like QMS or FTIR
|
|