TECHNICAL INFORMATION SHEET:
NEMOTO NP-18S Single Header Pellistor Gas Sensor
General Description
The Nemoto NP-18S is a catalytic (pellistor) type
flammable gas sensor supplied as a matched pair of
pellistor elements mounted on a single header and
protected by a metal mesh enclosure and can.
The sensor detects and measures the presence of
flammable gases and vapours in air, in the range 0-
100% of the Lower Explosive Limit (LEL) of the gas
or vapour being measured. Designed as a lower
cost alternative to the twin-header NP-18 device,
the NP-18S may be used as the sensing platform in
fixed flammable gas detection systems.
The NP-18S exhibits excellent long term zero and
sensitivity stability and a high level of resistance to
catalytic poisons. The highly automated
manufacturing procedure employed by Nemoto
results in a repeatable reliable sensor which, unlike
similar devices, requires no trimming resistor to
enable the detector to be matched with a
compensator.
Nemoto has a policy of continuous development and
improvement of its products. As such the specification for the
device outlined in the data sheet may be changed without notice
Specifications:
Recommended Voltage: 2.6V +/- 0.1V
Current Drawn: 170 +/- 20mA
Zero Offset: 0mV +/- 30mV
Minimum Sensitivity: 60-90mV/% CH4/Air
Range: 0-100% LEL
Accuracy: +/- 1%LEL(CH4)
Maximum Long Term Drift:
Span: < +/- 5% LEL/3 Months
Zero: <+/- ˝ mV/Month
Response Time: T50: 3 sec T90: 8 sec
Recommended Circuit:
Note: The value R1 is arbitrary, since the function of R1 is
to balance the bridge. 1K is suggested.
Temperature Range: -20°C to +70°C
Temperature Drift: (-20°C to +70°C)
Zero: < +/- 2%LEL
Humidity: 0-100%RH, non-condensing
Humidity Response: +/- 2%LEL
Linearity: Effectively Linear to 60%LEL
Test data on drift, poisoning, temperature
performance, linearity will be available on the
Characterisation Document.
ds-np18s.doc, issue 3, May 2008
|
|