PORTABLE STROBOSCOPES (external phase-shifting)

5

Phaser-Strobe allows three flash rates to be stored - displayed in flashes per minute orflashes per second.

Phaser-Strobe Series

NIST

Phaser-Strobe operates with internal rechargeable batteries or continuously fromAC line power with the optional power supply/recharger. Phaser-Strobes incorporate the unique design features of the Nova-Strobe DB Plus family with anincreased operating range of 30 to 32,500 flashes per minute, as well as external phase shifting. The unique digital adjustment knob can select the decade for adjustments, so coarse and fine adjustments of flash rates are made quickly and with significantly better resolution than competitive units. The memory feature of the Features N.I.S.T. Traceable Certificate of Calibration included • Phase Shift adjustable as phase angle or time with resolution to 0.1º and 0.0001sec • Auto step mode provides slow motion viewing for high speed events • Backlit alphanumeric LCD shows flash rate, degrees, time • Store and recall three memory settings • TTL compatible input/output jacks, power for optional sensors • Tachometer mode from external trigger Common Applications:• Calibration of Tachometers• Diagnostic Inspection• Engine R&D• Textiles Phaser-Strobe PB 115

SpecificationsPhaser-Strobe PBFlash Range 30-32,500 FPM (Flashes/Minute) 0.5-541.66 FPS (Flashes/Sec) (Hz) Accuracy ±0.01% of reading Digital Adjustment Knob Decade change push mode enables any value to be changed in decade (All ranges and modes) steps (power of 10) from 0.0001 per detent to 10,000 per detent Flash Rate Resolution ±0.05 FPM from 30 to 2,999 FPM, ±0.5 FPM from 3K to 32.5K FPM Flash Rate 5-32,500 FPM 0.5000-541.66 FPS 0-359 ±0.1100-3,499.0 FPM0-355 ±0.13,500-9,999.9 FPM0-350 ±0.110,000-32,500 FPM (External Triggering) Phase delay - Degrees
Phaser-Strobe PB Kit 230 Same as Phaser-Strobe PB Kit 115, but withPSC-4 (230 Vac) power supply/recharger
Time Delay - Seconds 0 to 6.500 seconds ±0.0001 seconds
Ordering InformationPhaser-Strobe PB 115 Phaser-Strobe PB 230 Same as Phaser-Strobe PB 115, but with R-6 (230 Vac) recharger
Auto Step Mode (Range) Step Size 0-180º Step Rate 0.0001-6.50 seconds per step
Phaser-Strobe PB Kit 115 Stroboscope,R-5 (115Vac) recharger Same as Abovewith PSC-3 (115 Vac) power supply/recharger and spare lamp in latching case
Flash Energy (Typical) 150 mJoule < 10,000 FPM, 50 mJoule > 10,000 FPM Flash Duration (Typical) 10-30 microseconds Average Power - Watts 10W Tachometer Mode 5-200,000 RPM from external trigger (flash inhibited above 33,000 FPM) External Input Input Pulse - 5 microseconds min, TTL to 24V max (1/8" phone plug) Trigger Output/Remote Sync 5V TTL compatible negative edge, 500 microseconds pulse (Typical) Power Internal rechargeable batteries or AC power supply/recharger Weight 2.5 Lbs. (1.2 kg) including batteries

PORTABLE STROBOSCOPES (for use with Vibration Data Collectors) Vibration-Strobe Series

Vibration Strobe Vibration Strobe is uniquely designed to provide precise, instantaneous synchronization to a num-ber of data collectors and Fft analyzers triggered by an accelerometer. Built for rugged, portable applications, the is the perfect lightweight phase analysis tool. Vibration Strobe allows for the measurement of phase without stopping the machinery to install reflective tape. Phase analysis is quick and accurate using the Filter Bandwidth Selector and the Relative Phase Adjustment. Unique “Tracking Filter” maintains phase lock to input pulse.Kits include: Strobe, interface cable, fast recharger and spare lamp in latching carrying case.

SpecificationsVibration StrobeFlash Range 100-12,000 Flashes Per Minute (FPM) Display 5-Digit Backlit Liquid Industrial crystal Display Accuracy ± 1 FPM Flash Energy 180 mJoule Flash Duration 30 µSec. Lamp Life (Typical) 100 million flashes External Trigger TTL (24V max) provides 5V out Tracking Filter Selectable Wide and Narrow Bandwidths Operating Time 60 minutes at 6,000 FPM Relative Phase Adjustment Selectable -90° to +90° Power 6 Vdc, Internal Rechargeable Batteries;115Vac, 50-400 Hzor 220-240 Vac, 50-400 Hz Fast Charger

Vibration Strobe VB-115

Ordering InformationContact Factory for available Models.
Weight 2.5 Lbs. (1.1 kg)