SuSix_US_datasheet_1202
6Pages

{{requestButtons}}

Catalog excerpts

SuSix_US_datasheet_1202 - 1

Data Sheet Turbidity and Suspended Solids Transmitter 1 Turbidity and Suspended Solids Transmitter As our products are continuously improved, we reserve the right 1 to make any change in the stated specifications and dimensions without prior notice.

Open the catalog to page 1
SuSix_US_datasheet_1202 - 2

Data Sheet Turbidity and Suspended Solids Transmitter MJK SuSix® turbidity and suspended solids sensors are designed for measurement by immersing the sensor into open tanks, wells and containers or in-line pipe mounting. The SuSix® transmitter is rugged and designed to handle tough applications. The scratch-proof saphire optics, the PUR cable and the heavy duty sensor housing ensure long life, even where scaling and bio-fouling are problems. Applications Sewage and wastewater plants Sludge treatment Application examples for measuring with SuSix® on a wastewater plant. Drinking water...

Open the catalog to page 2
SuSix_US_datasheet_1202 - 3

Data Sheet High Quality Materials Turbidity and Suspended Solids Transmitter The SuSix® sensor is constructed of stainless steel with chromium-dioxide coating and scratch resistant sapphire lenses in a highly polished stainless steel sensor face. An optional self-cleaning wiper system is available for particularly dirty and difficult applications. Flexible Installation MJK’s modular design allows up to 950 ft. between the sensor and the converter, and the Display Unit can be mounted up to 3000 ft. from the signal converter with ordinary twisted wires. One Display Unit can control up to four...

Open the catalog to page 3
SuSix_US_datasheet_1202 - 4

Data Sheet Specifications Turbidity and Suspended Solids Transmitter Converter Accuracy Measuring input Analog output One active 4 - 20 mA, galvanically isolated (max. load 800 W) Digital outputs One voltage-free electromechanical relay (max. 50 V DC / 1 A) One optically isolated MOSFET relay (max. 50 VAC / V DC / 120 mA) Digital inputs One, max. 30 V DC, < 5 V DC = 0 (low), > 10 V DC = 1 (high), pulse length > 100 ms MODBUS® RTU-mode, 9600 baud, 2-wire RS 485, slave-mode RS 485 for connection to Display Unit or PLC Power supply Cabinet material 24 V AC, 50 / 60 Hz ± 10 % or 115 V AC, 50 /...

Open the catalog to page 4
SuSix_US_datasheet_1202 - 5

Data Sheet Specifications Turbidity and Suspended Solids Transmitter SuSix® Sensor Turbidity Suspended solids Measurement Principe Infrared-diode system and beam focusing (l = 860 nm) Two-channel-90° scattered light measuring corresponding to EN 27027/ISO7027 Additional verification by six-channel multi-angle measuring Six-channel, multi-angle with modified absorption 2 channels are used for absorption Suspended solids Materials Head: Stainless steel DIN 1.4460 — Body: Stainless steel DIN 1.4435 Optical lenses: Sapphire — O-ring: Viton Wiper (optional): PA (GF), TPU — Cable: PUR 3 x 2 x 20...

Open the catalog to page 5
SuSix_US_datasheet_1202 - 6

Data Sheet Turbidity and Suspended Solids Transmitter Mechanical Dimensions Converter and Display Unit Punch holes for 3/4” conduit connectors Order Numbers Max. four MJK converters with Modbus can be connected to one display unit: Display unit 4 wire shielded cable. display unit 2 wire shielded cable. B Fuse 24V AC: 500mA T Fuse 115V AC: 125mA T Fuse 230V AC: 63mA T display unit display unit Max. four MJK converters with Modbus can be connected to one display unit: Display unit 4 wire shielded cable. 2 wire shielded cable. B Fuse 24V AC: 500mA T Fuse 115V AC: 125mA T Fuse 230V AC: 63mA T...