1
PUMP % Flow rate DISP o Volume DISP o Time PAUSE o Time Total o
Fließrate kalibrieren
Calibrating the flow rate
Calibration du volume
1 Mit der MODE-Taste auf PUMP • Flow rate wechseln. Mit den at Tasten die ge­wünschte Fließrate eingeben k Kontrolle: richtiger Schlauch-i0 gewählt? (siehe Seite 15)
2 MitderMODE-Tasteauf DISP • Time wechseln
- Mit den at Tasten 60 Sek. eingeben
- Pumpe mit RUN/STOP-Taste starten
k Pumpe stoppt automatisch
nach 60 Sekunden
k Dosierte Flüssigkeit nach
Volumen oder Gewicht
bestimmen.
Erhaltenen Wert wie folgt kalib­rieren:
3 MitderMODE-Tasteauf PUMP • Flow rate wechseln.
4 CAL/MAX-Tastedrücken (Anzeige blinkt)
Mit den at Tasten gemessenen Wert eingeben (Waage)
5 Mit der CAL/MAX-Taste spei­chern (die Fließrate wird dem vorgegebenen Sollwert best­möglich angepasst).
Die Drehzahl hat eine Auflö­sung von V100 min^1. Nach der Kalibrierung kann die ursprünglich eingegebene Fließrate daher oft nur ange­nähert werden.
2
Passer avec la touche MODE sur PUMP • Flow rate Introduire le débit souhaité au moyen des touches at k Contrôler si le 0 int. de tube adéquat a été choisi (voir page 15).
Passer avec la touche MODE sur DISP • Time
- Introduire 60 secondes avec les touches at
- MisenrouteavecRUN/STOP k La pompe s'arrête automa­tiquement après 60 secondes k Déterminer le liquide dosé selon le volume ou le poids.
2
Change mode to
PUMP • Flow rate
Enter the required flow rate
with the at keys
k Check whether the correct
tubing i.d. is selected
(see page 15).
Change mode to DISP • Time
- Enter 60 seconds by using the at keys
- Start pump with RUN/STOP k Pump stops automatically after 60 seconds
k Measure the dispensed li­quid by volume or weight.
3h.O,
PUMP o Flow rate DISP o Volume DISP -fcTime PAUSE o Time Total o
2
default
MODE
bO.O
PUMP -fc Flow rate DISP o Volume DISP o Time PAUSE o Time Total o
3
Calibrate the ascertained value as follows:
3 Return to mode PUMP • Flow rate.
4 PresstheCAL/MAX-button (displayed value blinks) Enter the measured value (ba­lance) with the at keys
5 ConfirmwiththeCAL/MAX-key (the flow rate setting is re­adjusted to the initially entered set point as close as possible). The rotation speed has a reso­lution of V100 rpm.
After calibration, the initially set flow rate can only be approached.
Calibrer la valeur ainsi obtenue comme suit:
3 Passer avec la touche MODE sur PUMP • Flow rate.
4 PresserlatoucheCAL/MAX (l'affichage clignote) Saisir la valeur mesurée (balance) avec les touches at
5 Mémoriser avec la touche CAL/ MAX (le débit retourne main­tenant automatiquement dans la zone de la valeur préréglée Initialement). La résolution de la vitesse de rotation est de Vu» t/min. Après la calibration, Il arrive assez souvent que le débit défini initialement ne puisse seulement être approché.
k Calibration par défaut: voir page 22
4
ok
CAL MAX
5
OL n -IÖ.U
k Default-Kalibration setzen siehe Seite 22
k Setting the default calibration see page 22