Metrosep A Supp 4 – 250 (6.1006.430)
The Metrosep A Supp 4 is an extremely robustcolumn with very good separation properties. The
separation phase consists of polyvinyl alcohol particles
with a diameter of 9 µm, which are functionalized
with quaternary ammonium groups. This arrangement
guarantees a high stability and great tolerance to
very fine particles that could pass through the built-
in filter plate. The Metrosep A Supp 4 has a medium
ion exchange capacity; sulfate elutes after 12.5 min.
The number of plates that can be achieved with this
column is higher than with the Metrosep Dual 2
separation column. For 5 mg/L chloride and 10 mg/L
sulfate, 5900 respectively 5800 theoretical plates
have been determined. Metrosep A Supp 4 is parti-
cularly suitable for all routine tasks in
water analysis.
Its universal applicability makes it into the standard
column for the Compact Metrohm IC Systems 792
Basic IC, 761 Compact IC and 861 Advanced
Compact IC. Technical information Carrier materialPolyvinyl alcohol with quaternaryammonium groupsColumn dimensions250 x 4.0 mm
Column bodyPEEK
Standard flow1.0 mL/min
Maximum flow2.0 mL/min
Maximum pressure12 MPa
Particle size9 µm
Organic modifier0...100%
pH range3...12
Temperature range20...60 °C
Capacity71 µmol (Cl Applications • standard anions
• water analysis
• method development
• difficult matrices
• critical samples
• iodide >
– ) Eluents Without chemical suppression
Phthalic acid eluentPhthalic acid1660 mg/2 L5.0 mmol/L
(standard eluent)Acetone40 mL/2 L2.0%TRISpH = 4.4With chemical suppressionCarbonate eluent Sodium hydrogen carbonate286 mg/2 L1.7 mmol/L
(standard eluent)Sodium carbonate382 mg/2 L 1.8 mmol/LCarbonate eluent, mod. 1Sodium hydrogen carbonate336 mg/2 L2.0 mmol/LSodium carbonate509 mg/2 L 2.4 mmol/L
Acetone100 mL/2 L5.0%Carbonate eluent, mod. 2Sodium hydrogen carbonate168 mg/2 L1.0 mmol/L Sodium carbonate848 mg/2 L 4.0 mmol/L Ordering information 6.1006.430Metrosep A Supp 4
6.1006.500Metrosep A Supp 4/5 Guard
6.1006.540Metrosep A Supp 4/5 S Guard
To protect the IC column – even though it is notparticularly sensitive – we recommend the use of the
Metrosep A Supp 4/5 Guard column (6.1006.500).
If the column oven is used then Metrosep A Supp
4/5 S Guard column (6.1006.540) must be used. 50 >