Text version of the page
IC Columns
Our range of
Care Chromatograms: Metrosep Anion Dual 2 Regeneration:Rinse with 0.1 mol/L HNO
3 at 0.3 mL/min forapprox. 2 h; then switch to standard eluent.
Rinse with an eluent containing 100 mmol/L of
the particular buffer salt.Organic contaminants:Rinse the column with eluent containing organic
solvent (max. 20%).Storage:In the eluent
Phthalic acid eluent, standard 1Fluoride5.00 mg/L
2Chloride5.00 mg/L
3Nitrite5.00 mg/L
4Bromide10.00 mg/L5Nitrate10.00 mg/L6Sulfate10.00 mg/L
7System peak Phthalic acid eluent, drinking water1Chloride5.23 mg/L2Nitrate10.32 mg/LCarbonate eluent, standard1Fluoride5.00 mg/L2Chloride5.00 mg/L
3Nitrite10.00 mg/L
4Bromide10.00 mg/L5Nitrate10.00 mg/L6Phosphate10.00 mg/L
7Sulfate10.00 mg/L3Sulfate6.94 mg/L4System peak Carbonate eluent, drinking water
1Fluoride0.05mg/L
2Chloride4.73mg/L
3System peak4Nitrate10.06mg/L5Sulfate6.20mg/L Carbonate eluent, standard
1Chlorite,ClO >
2– 40.00mg/L2Chloride, Cl >
3– 2.00mg/L4Perchlorate, ClO >
– 4.00mg/L3Chlorate, ClO >
4– 1.00mg/L >
41