IC Columns
Our range of
Care Regeneration:Light contamination:
Rinse with NaOH/NaAc (100/500 mmol/L)
at a flow rate of 1 mL/min for 4 h, followed
by the eluent at the same flow rate for 12 hHeavy contamination:Rinse with 100 mL eluent with 50% acetonitrile
at a flow rate of 1 mL/min.Storage:In the eluent Remark Start the system with a flow rate of 0.3 mL/min.
Adjust to the desired flow rate after 10 min.
Likewise reduce the flow rate of the pump to
0.3 mL/min for 10 min before turning off . Chromatograms: Metrosep A Supp 8 – 150 >
Sodium chloride eluent, standard, UV detection at 215 nm 1Nitrite1.00 mg/L
2Bromide20.00 mg/L
3Nitrate2.00 mg/LSodium chloride eluent, standard in 100 g/L NaCl, UV detectionat 215 nm1Nitrite0.05 mg/L2Bromide1.40 mg/L3Carbonate-4Nitrate0.12 mg/L Ordering information 6.1017.010Metrosep A Supp 8 – 150
6.1017.500Metrosep A Supp 8 Guard >
Cyanide eluent, standard
1Sulfide20.0 µg/L
2Cyanide20.0 µg/LChloride elutes after approx. 21 min. 37 >